background image

Reference Manual 
 
Power Analyzer EMM5 

 

Rev. 11 

 

 

2018-07 

 

 

2

 

Measurement system 

The EMM5 is designed to acquire all necessary data to provide full information in a 3-phase power distribution system. 

This requires the measurement of the three voltages L1/2/3 versus the N connector and the three currents in L1/L2/L3. 

Additionally, the current in N can be measured or, if desired, it can also be calculated from the data for I1/I2/I3 (but this 

will be less accurate due to calculation and rounding errors).  

The  EMM5  provides  separate  input  terminals  for  device  supply  and  measurement,  so  the  measurement  inputs  are 

completely independent. Of course, measurement and supply may be connected to the same power grid. 

For the most measurement values, minimum and maximum is stored and they are available for the user if necessary. 

2.1

 

Data collection  

The waveforms of the signals (=voltage or current) on the 6 (7 with N-current) input channels are sampled by the data 

collection logic of the EMM5. This supplies the software with enough information to calculate all the values.  

The sampling of the input signal needs to be synchronized to the input signal, so the EMM5 needs at least one input 

signal for the voltage L1-N to be able to do calculations. If this voltage is too small or even disconnected, the device 

will not be able to do any measurements. This mechanism is also used to provide the frequency reading. 

The data collection circuitry for the voltage channels uses the four terminals L1, L2, L3 and N which have 

to be connected in accordance to the connection diagram. As a minimum, the L1/L2/L3 terminals have to 

be connected while N can be left floating. In this case, N is artificially generated inside the device by three 

resistors. Connection of the N-terminal overrides this high-impedance resistor circuit. The voltages can be 

connected directly (with respect to the maximum rating) or via transformers. The voltage transformer factor can be set 

in the EMM5. 

The  three  (four  with  N)  current  inputs  use  two  terminals  each,  labelled  “K”  and  “L”.  The  neutral  current  can  be 

measured, but if necessary it can also be calculated with the other 3 currents. Current transformers have to be used for 

the measurement. The current transformer factors can be set in the EMM5. Each channel is additionally equipped with 

an input filter. This protects the system from being damaged by transients with high frequencies like voltage spikes and 

is mainly used to ensure proper sampling, synchronisation and measurement in presence of high frequency harmonics.  

Always make sure, you don’t exceed the maximum ratings of voltage and current channels.  The ratings 

and a connection diagram are available on the labels on the back side of the device. 

 

Summary of Contents for EMM5

Page 1: ...nual Power Analyzer EMM5 Rev 11 2018 07 Beluk GmbH Taubenstrasse 1 86956 Schongau Germany Tel 49 0 8861 2332 0 Fax 49 0 8861 2332 22 E Mail blr beluk de Web http www beluk de Reference Manual POWER AN...

Page 2: ...updates 14 12 04 CE 03 features of new software V1 05 18 03 05 ATh 04 features of new software V1 06 08 12 05 ATh 05 features of new software V1 07 connection diagrams added 28 02 06 ATh 06 features...

Page 3: ...3 phase measurement with neutral conductor and four current transformers four wire measurement 19 5 2 3 phase measurement without neutral conductor with three current and three voltage transformers t...

Page 4: ...erformed on the input data from all channels 3 voltage channels 4 current channels This provides information about the harmonic contents which distort the sine waveforms of currents and voltages The f...

Page 5: ...o small or even disconnected the device will not be able to do any measurements This mechanism is also used to provide the frequency reading The data collection circuitry for the voltage channels uses...

Page 6: ...hese further calculations are possible THD U THD for each voltage channel from the harmonics 1 62 THD I THD for each current channel from the harmonics 1 62 Pth Ith Exponentially damped values for cur...

Page 7: ...alarms Each of them continuously compares one of the measured values to an assigned limit and if necessary triggers the alarm Each of the alarms provides the following features Alarm source can be an...

Page 8: ...the reverse display bright number on dark ground shows an activated relay The message ALARM xx in the centre of the display shows the number of the active alarm The number appears instead of xx in thi...

Page 9: ...ntrast set up menu in this case so an additional possibility to set the contrast is included in the introduction screen press one of the two keys in the middle repeatedly to change contrast If no chan...

Page 10: ...lue becomes too big the display automatically switches the unit prefixes For example If the voltage exceeds 1000V the display switches to 1 00 kV In the top right corner of the display the status of a...

Page 11: ...rmonic alarms should use the THD as source which contains information of all harmonics of one source in one single value 4 5 Work This submenu contains the work accumulators These accumulators count c...

Page 12: ...ible settings the setup menu is divided into more submenus to provide easy and logical access to all the set up possibilities The available submenus are PARAMETER This menu provides the possibility to...

Page 13: ...te it by pressing the key PT RATIO This sets up the device to work with a voltage measurement transformer Enter the transformer ratio The range is 1 4000 If no transformer is used and the EMM5 is dire...

Page 14: ...are of reactive nature reactive power cp the display shows ind or cap for the trigger setting This has to be interpreted as more inductive than limit or more capacitive than limit LIMIT The limit val...

Page 15: ...lose if P sum 20W relay 2 will close when P sum 20W This setting contains the dead zone of 20W The on delay of 1sec and off delay of 0 5 sec for both relays prohibits them from being closed at the sam...

Page 16: ...lse outputs Select one pulse output with the and keys and after pushing the value can be adjusted The range is between 1 and 10000 pulses Mwh or Mvarh The adjusted value should be selected in the way...

Page 17: ...ry For the DI can be adjusted if the synchronization is done via a positive edge HIGH or negative edge LOW When synchronization via the DI the time interval to be terminated since this time will be us...

Page 18: ...torage cells full the recording starts again at memory cell one As additional information will be stored the values listed in the following table Page 1 Page 2 Page 3 Record 1 64 Indicates in which me...

Page 19: ...wer systems with or without the neutral conductor Below measurements in systems with L1 L2 L3 and N PEN are called four wire measurement Measurements in systems without N are called three wire measure...

Page 20: ...Rev 11 2018 07 20 5 2 3 phase measurement without neutral conductor with three current and three voltage transformers three wire measurement 5 3 3 phase measurement with two current and three voltage...

Page 21: ...nalyzer EMM5 Rev 11 2018 07 21 5 4 3 phase measurement with two current transformers three wire measurement 5 5 3 phase measurement with three current transformers and measured neutral conductor curre...

Page 22: ...ltage max 250V DC switching current max 100mA switching frequency max 4Hz tON 50ms tOFF 50ms Digital Input On request Fan control temperature measurement on rear side of device programming of relay ou...

Reviews: