Chapter 5: Technical Specifications
45
ENVIRONMENTAL
Operating Temperature
10ºC to 32ºC (50ºF to 90ºF)
Storage Temperature
-15ºC to 40ºC
Relative Humidity
10% to 90%
Pressure
49-103 kPa
Shock and Vibration
Meet MIL STD.810E method 514.4 (Basic
Transportation)
OPERATING
PARAMETERS
Flow Rate
10 -750 ml/min, with a 1000 ml/min as an option, in 10
ml/min steps plus 2.5 and 5.0ml/min with fluids of
viscosity 1 to 8 centipoise (Water and crystalloid fluids
through packed red cells)
Tolerance:
± 10% from 20 - 1000 ml/min
± 25% for 2.5, 5.0,10 ml/min
Output Temperature
Set to 37.5ºC for flow
≥ 60ml/min, to 39ºC at 50ml/min
or lower.
Tolerance: 1ºC for fluid temperature between 30ºC to
40ºC and 2ºC outside this range
Heating Capacity
Min. 1400 watts to fluid (20ºC temperature rise at
1000 ml/min)
Line Pressure
0 - 300 mm Hg, via pressure transducer
Operating Modes
a)
Load disposable set
b)
Prime system
c)
Prime patient line
d)
Infuse at operator controlled rate with warming
e)
Infuse fixed volume bolus with warming
f)
Stop system