Chapter 2: Operation
6
INSTALLING DISPOSABLE SET
WARNING:
The disposable set is for single patient
use only. Do not reuse.
Store the disposable set in a dry well-
ventilated area free from exposure to
chemical vapors.
We strongly recommend loading and
priming the disposable set just prior to the
procedure.
3-Spike Disposable set with key components
1. Snap reservoir
chamber into the
reservoir support
clamp.
2. Open the door. Insert
heat exchanger with
red arrow pointing up
(Red tinted tubing
to
red stripe on
unit.)
3. Firmly position the
interlock block into the
fluid out detector.
4. Guide the curved piece
of pump tubing (
Blue
tinted tubing
) over the
pump head. Check that
the thinner recirculate
line is in the grove to
the right.
Do not kink or twist
the tubing
5. Place
the
pressure
chamber
into
the
pressure chamber well.
Firmly insert the wider
infuse line into the air
detector and to the left
of valve wand.
Do not apply excessive
pressure to the
pressure transducer.
The pressure
transducer can be
damaged with
excessive force. Do not
use the system if the
pressure transducer is
damaged.
6.
Place the thinner recirculate line to the right of
the air detector, and to the right of the valve
wand.
7.
Close and latch the door. Make certain the
pump tubing is not caught. Connect the patient
line.
Check Valve
Roller Clamp
Infuse
Line
Bag Clamps
Air Vent
Reservoir
Chamber
Interlock Block
Pump Tubing
Recirculate
Line
Patient
Line
Pressure
Chamber
Red Tinted
Tubing
Coarse
Filter
Heat Exchanger
Bag Spikes