Chapter 4: Parameters Setting and Preventative Maintenance
33
5.
Hardware Verification
Install and prime the disposable set before beginning the Hardware Verification process.
Hardware
mode verifies:
a.
Valve operation
b.
Fluid Out and Air Detectors
c.
Battery voltage.
d.
Flow Rate (Pump speed)
e.
Input and Output Temperature Probes
including “Over Temperature”
alarm test, and
f.
Pressure sensor.
A password is required to access the SERVICE screen, to insure that this mode is not
accessed accidentally.
Press the SERVICE key, at power-up, to access the Calibration/Set-up screen. This
screen remains active for 4.5 seconds before the system enters the Prime mode screen.
WARNING!
Do not access hardware verification while the instrument is patient
connected.
Press HARDWARE from the Calibration/Set-Up screen.
Enter the Password 013192.
Calibration/Setup Screen
CALIBRATION/SET-UP
TIME 23:59 DATE: 12-31-14
BOLUS 200 ML PRESS LIMIT 300MMHG
AC POWER PRESENT
DATE
TIME
DISPLAY
BRIGHT
TEMP
CAL
PRESS
CAL
POWER
CAL
MEDIUM
KEYRATE
PRESS
LIMIT
HARD-
WARE
SETUP
BOLUS
EXIT
SERVICE