Chapter 5: Technical Specifications
5-2
ENVIRONMENTAL
Operating Temperature
10
o
C to 32
o
C (50
o
F to 90
o
F)
Storage Temperature
-15
o
C to 40
o
C
Relative Humidity
10% to 90%
Pressure
49-103 kPa
Shock and Vibration
Meet MIL STD.810E method 514.4 (Basic
Transportation)
Electromagnetic
Compliance
Meet EN60601-1-2 (2007) and IEC 60601-1-2
(2007)
OPERATING
PARAMETERS
Flow Rate
10 -1000 ml/min in 10 ml/min steps
Tolerance:
±
10% from 20 - 1000 ml/min
±
25% for 10 ml/min
Output Temperature
User adjustable for target temperature of 37
o
C to
48
o
C
Heating Capacity
Min. 1400 watts to fluid (20
o
C temperature rise at
1000 ml/min)
Line Pressure
0 - 300 mmHg, via pressure transducer
OPERATING PANEL
Control Panel and Display
Splash proof touch screen display
Display Area
Diagonal screen 5.7” (14.5 cm)
Status Display
Flow rate (ml/min)
Total volume infused (ml)
Line pressure (mmHg)
Output fluid temperature, Tpump (
o
C)
Patient temperature @ location 1, T1 (
o
C)
Patient temperature @location 2, T2 (
o
C)
Patient temperature @location 3, T3 (
o
C)
Patient temperature @location 4, T4 (
o
C)
Target temperature (
o
C)
Alarm messages
Functional Keys
Keys are displayed appropriate to the particular
point in operation
Character Display
Graphical Alarm Messages - display where errors
have occurred