Chapter 3: Alarms and Troubleshooting
3-2
DOOR OPEN
Door Open Alarm Screen
The door is open.
No magnet in the
door latch.
Close the door to silence the alarm
and resume.
Check magnet in the door latch.
If the door is opened while the system
is pumping, the system will
immediately stop heating and
pumping. The valve moves to the
recirculate position and an audible
alarm sounds.
FLUID OUT
Fluid Out Alarm Screen
Fluid Out Message after Pressing
REPRIME Screen
Out of fluid.
Bag clamps not
fully opened or
fully spiked.
Tubing in the Fluid
out sensor is not
seated firmly in
the detector, or
tubing is stretched
or pulls away from
the sensor, due to
vacuum in the
line.
Clogged filter.
Reservoir or
recirculate line is
obstructed.
Vacuum source is
set too high.
Detector
electronics
defective.
Press MUTE to silence the alarm.
If out of fluid, add additional fluid and
press REPRIME.
If the Reprime volume count does not
count down from 100 to 0 ml, then:
• Check the bags are fully spiked
and clamps are fully opened.
• Check that the pump head tubing
is not stretched and is seated
firmly within the fluid out sensor.
• Check the fluid out sensor and
make certain it is clean and there
is nothing obstructing contact with
the sensor.
• If there is fluid in the reservoir,
check the vacuum source.
Vacuum should not be more than
100 mmHg.
Reseat the tubing in the fluid out
detector and make certain that it is
seated firmly in the sensor.
High amounts of particulates may
clog the coarse filter in the reservoir.
Replace the reservoir if it is clogged.
Power down and service the machine
if error persists.