PN#P5SP013 | Rev. A | International
15
14
PN#P5SP013 | Rev. A | International
Battery Charge Indicator in Charging state:
After 30 minutes of being off the charger the SkinPen
®
Precision will beep
to remind users that it needs to be placed on the charger. This notification may
be cleared by either simply powering on the device, or by placing the device on
the charger.
•
When the SkinPen
®
Precision is placed on the charger, LED
1, 2, 3 and charger indicator LED will blink sequentially for a
second to indicate it is connected with the charger base.
• When the SkinPen
®
Precision is removed from the charger
base the LED's stay solid which indicates the charge level in
the battery. They then go off to maintain battery longevity.
• Battery Charge = 100%:
LED 1, 2, 3 OFF to maintain battery longevity.
• Battery Charge > 90%:
LED 1, 2, 3 ON.
• 70% < Battery Charge ≤ 90%:
LED 1, 2 ON, LED 3 repeat on/off 1 sec.
• 30% < Battery Charge ≤ 70%:
LED 1 ON, LED 2 repeat on/off 1 sec.
• Battery Charge ≤ 30%:
LED 1 repeat on/off 1 sec.
13. SPECIFICATIONS
Technical Information of SkinPen Precision
®
Product Name
SkinPen Precision
®
SkinPen
®
Precision Handpiece
Model Number
100
SkinPen
®
Precision Charger Base
Model Number
101
Bellus Aesthetics FDA Registration #
3010392991
FCC ID
2AGLK-101
Weight and Unit
≤ 5oz /155mm length and
max. outer diameter of 34mm
Electrical Requirements
Charger Base Input: 5VDC, 2A max
Output voltage:
5W (max)
Charger Time
From 10% charge to 90%
Charge within 14 hours
Working Time
> 6 hours
(under normal use conditions)
Speed
6300RPM – 7700RPM
Needles
•
14 total solid needles
•
32 BWG (gauge)
•
<32 RMS (roughness)
•
Medical grade Stainless Steel
•
EU RoHS compliant
•
Sharpness specification within the
Radius 0.005mm (Max)
•
Maximum extension of the needles
from the needle head surface is less
than 2.7mm
Operation Cordless
AC Adapter
Medical Grade, Universally
compatible power requirements:
100-240 VAC at 50-60 Hz
Charger base transmitting frequency
Between 110kHz and 205kHz
with efficiency around 73%
70% <
Battery
Charge
3
2
1
3
2
1
Battery
Charge ≤ 30%
Battery
Charge = 100%
3
2
1
3
2
1
30% <
Battery
Charge
3
2
1
3
2
1