How to charge:
• Inductive charging is used between the SkinPen
®
Precision
charger base and the SkinPen
®
Precision device.
• Plug the charger base into a live outlet.
• Place the hand-piece into the base with the power button
facing out. See “FAQ/Troubleshooting” for additional
battery information. Battery charge percentages in “FAQ/
Troubleshooting”.
Power:
* Powering ON/OFF should only be done with the SkinPen
®
Precision
device disconnected from the charging base.
• ON: Press and hold power button for 1 seconds.
• OFF: Press and hold power button for 0.5 seconds.
8. PROCEDURE INSTRUCTIONS, POST-PROCEDURE
INSTRUCTIONS, POST-PROCEDURE CARE
• For additional information refer to SkinPen
®
Precision IFU.
9. CLEANING OF SKINPEN
®
PRECISION AND CHARGER BASE
* Ensure SkinPen
®
Precision device is powered down before cleaning,
and SkinPen
®
Precision charger base is unplugged.
• The device should be cleaned while holding the SkinPen
®
Precision facing
straight down while wiping the rotary area. Do not clean near the seal.
• Germicidal disposable wipes should be used to clean the SkinPen
®
Precision
after each procedure. Germicidal disposable wipes may also be used to clean
the SkinPen
®
Precision Charger Base. Germicidal disposable wipes should be
used to carefully wipe the SkinPen
®
Precision according to their directions
for use found on the Germicidal disposable wipes labeling. Attention should
be paid to clean areas such as crevices, seams, and areas around where the
SkinPen Precision Cartridge attaches to the device.
• Germicidal disposable wipes that protect against HIV-1 and HBV of surface/
objects soiled with blood/body fluids are recommended to be used.
DIRECTIONS FOR USE:
Special instructions for cleaning & decontamination against HIV-1 and HBV of
surfaces/objects soiled with blood/body fluids:
Personal protection:
Specific barrier protection items to be used when handling
items soiled with blood or body fluids are disposable latex gloves, gowns,
masks, or eye coverings.
Cleaning procedure:
Blood and other body fluids must be thoroughly cleaned
from surfaces and objects before application of the disinfectant.
Disposal of infectious materials:
Blood and other body fluids should be
autoclaved and disposed of according to federal, state and local regulations for
infectious waste disposal.
Contact time:
Leave surfaces wet for 30 seconds and 10 minutes for HIV-1 and
HBV, respectively. Use the 10 minute contact time to mitigate other viruses,
bacteria and fungi listed on the label.
• Do not immerse in liquids.
• Do not use solvents to clean device.
10. STORAGE
• For optimal performance of your SkinPen Precision
®
, ensure the device is turned
off and store the device in the SkinPen
®
Precision charging base when not in use.
• If the device is OFF and not connected to the charging base for 30 minutes,
the device will generate a “Not on Charger” alert by “beeping” repeatedly at 1
second intervals for 1 minute, and then every 10 minutes after as a reminder
to return the SkinPen
®
Precision to the charger base. Turn off the alert by
connecting the SkinPen
®
Precision to the charger base.
11. DISPOSAL
• Dispose of cartridges/needle tips as medical waste via a Sharps container.
• Properly dispose of all items in accordance with local regulations.
• You must dispose of SkinPen Precision
®
, SkinPen
®
Precision Charger, and all
other SkinPen® Precision components properly according to local laws and
regulations. Because SkinPen
®
Precision contains electronic components and
a Lithium Ion rechargeable battery, SkinPen
®
Precision must be disposed of
separately from household waste. When SkinPen
®
Precision reaches its end of
life, contact local authorities for proper disposal and recycling options.