Technical Manual WD 230 and WD 250
Rev. 1.8 PF
Page 31
73018_E rev 1 8-Technical Manual.doc
4.6.6 Automatic Program Recognition
ON / OFF / Only, Default ON
If Reed sensor basket recognition S105 is closed, basket coding S
106 – 108 is read into memory. Depending on magnet combination,
some 7 different programs may be activated. Upon activating door
key, the automat will start the selected program after door closing.
S10
5
S10
6
S10
7
S108 Program
X X
P1
X X
P2
X X X
P3
X X
P4
X X X
P5
X X X
P6
X X X X
P7
S105 = Basket positioning
Basket Coding
On / Off / Only
ON
Program selection with
basket coding is active.
The entered programs
may be
overwritten by
means of the program
keys
.
OFF
Basket coding is not
active, programs may
activated by
keyboard
only.
Only
Program selection
only
by means of
basket
coding
. When pressing
program key, display
shows „only by basket
coding “.
4.6.7 Display and Message in case of error.
e.g. „Tel No Tech Service 105“
Error Message
Tech. Service 105
Assignment of Printer
.
Tabletop printer / Panel printer
Printer
Desk (Star SP200)