107
Door does not
close
If the closure process with US motor
(M38) takes longer than 16 s until the
upper switch (S2/K2) closes or the
potentiometer value (R2) is closed, the
controller terminates the process with
error display.
Input (S111), (R2) is monitored during the
closure.
S111
Door drive | Door belts:
US: M35 & REL35 open & REL36 close
CS: M37 & REL37 open & REL38 close
Motor doesn’t drive:
Door drive belts intact? Replace belts.
Contactor activates:
Check 3Ø voltage contactor, Check fuses
F11, F12, Check door motor capacitor.
Contactor doesn’t activate:
Check 24Vac, Replace contactor,
Check output LED’s on Control PCB
: Check wiring, Replace
Control PCB,
Motor drives:
Check upper/lower limit switches. Check wiring.
108
Object in
Door
US Safety switch actuator (S210) has
been triggered. If during the closure
process the safety switch actuator opens
Relay ESR1, the controller terminates the
closure process with error display and
reopens the door (M38). Input (S210) is
monitored only during the closure
process.
S210
Door tape safety switch
DS: ESR-1 & K1 CS: ESR-2 & K2
Obstruction in door track:
Clear obstruction.
Check Tape Switch Relay PCB:
Red LED: tape switch
disconnected, check wiring/plug, replace tape switch. Yellow LED:
tape switch tripped: No obstruction, replace switch.
Green LED: tape switch normal.
109
No
Communicati
on
Communication between processor and
door panel PCB (IPD) has been
disrupted.
110
No Pressure
Pressure in washing system has dropped
below minimum. If Switch (S211) is still
open during the program 90 s after the
filling phase, the controller terminates the
current program with error display.
Input (S211) is monitored only with
Utilities (M1-M4), (M7-M9), 90 s after the
program time and below 85°C. Monitoring
is inactive above 85°C.
S211
Pressure Switch S211:
Wash pump circulates water. Level sensor
ok:
Check for activation of S211. Replace switch.
Summary of Contents for WD 290
Page 1: ...Washer Disinfector Service Training Seminar Belimed US Charleston SC...
Page 3: ...Automatic vertical glass door Door opens down Use the same rack as the 230 WD 250...
Page 4: ...WD 290 15 tray washer disinfector with automatic glass sliding doors...
Page 5: ...Accessories instrument racks...
Page 8: ...4 key factors for cleaning efficacy Chemicals...
Page 19: ...TOSI testing...
Page 27: ...Dynamic fill advantages...
Page 28: ...WD 290 high voltage section Power Supply Circuit breakers Contactors...
Page 73: ...Drying system Separation of air and water systems...
Page 82: ...Service mode...
Page 83: ...Analysis...
Page 84: ...Dosing...
Page 85: ...Controls...
Page 86: ...Parameter 1 Analysis 2 Dosing 3 Controls 4 Parameter 5 Configuration 6 Limits...
Page 87: ...Program Structure...
Page 88: ...Configuration...
Page 89: ...Limits...
Page 90: ...WD 250 magnet template...
Page 96: ...Process status display...
Page 97: ...Menu...
Page 98: ...Log on...
Page 99: ...Password entry...
Page 100: ...Data entry example...
Page 101: ...Machine information...
Page 102: ...Error messages...
Page 103: ...Settings...
Page 104: ...Measurements...
Page 105: ...System configuration...
Page 106: ...Parameters...
Page 107: ...Function medium...
Page 108: ...Door position...
Page 109: ...Manual function...
Page 110: ...Magnet positions...
Page 125: ...PM check sheet...
Page 126: ...Wash are disassembly...
Page 127: ...Lower wash arm docking port...
Page 128: ...Upper wash arm...
Page 129: ...Daily screen cleaning...
Page 130: ...Wash Arm...
Page 131: ...Water supply strainers...
Page 132: ...Air HEPA pre filter...
Page 133: ...Drying system...
Page 134: ...Air water check valve...
Page 135: ...Air water check valve...
Page 136: ...Light bulb...