121
No Water
Level sensor is defective or not
connected.
If the level sensor (P1) is not closed for
longer than 2 s (value between 0.2V and
4.5V), the controller terminates the
current program
with error display. The threshold values
for Input (P1) are monitored during the
program cycle.
Drain valve
Drain valve not closing. If possible, check floor drain for water flow
during wash. If drain boot suspected, do not dismantle drain valve
until replacement part available.
Replace drain boot.
Level sensor
Faulty level sensor. Check voltage at transducer. Empty level:
0.5Vdc, Working level: minimum 1.0Vdc, dynamic reading.
Replace sensor | Replace sensor board | Check wiring | Check power
supply | Check input to control board | Replace control board.
130
NTC 1
Defective
The temperature sensor in the washing
chamber floor is defective or not
connected (value between 0°C and
150°C). The controller terminates the
current program with error display.
Drain valve
Drain valve not closing. If possible, check floor drain for water flow
during wash. If drain boot suspected, do not dismantle drain valve
until replacement part available.
Replace drain boot.
Level sensor
Faulty level sensor. Check voltage at transducer. Empty level:
0.5Vdc, Working level: minimum 1.0Vdc, dynamic reading.
Replace sensor | Replace sensor board | Check wiring | Check power
supply | Check input to control board | Replace control board.
131
NTC 2
Defective
Dryer temperature sensor (NTC2) is
defective or not connected. Cause and
remedy see Error 130
Temperature Sensors:
Blowers/heaters activate; no change in
temperature display:
Check wiring to sensors. Replace sensors. Check Control PCB. Error
code 132: may indicate inadequate air flow to sensor #2 in heater
compartment. Check that diffuser plate in heater compartment is
aligned to upper port.
133
NTC 3
Defective
Temperature sensor (NTC3) is defective
or not connected. Cause and remedy see
Error 130
134
NTC 4
Defective
PW tank temperature sensor (NTC4) is
defective or not connected. Cause and
remedy see Error 130
135
NTC 5
Defective
Washing chamber floor temperature
sensor (NTC5) is defective or not
connected. Cause and remedy see Error
130
Summary of Contents for WD 290
Page 1: ...Washer Disinfector Service Training Seminar Belimed US Charleston SC...
Page 3: ...Automatic vertical glass door Door opens down Use the same rack as the 230 WD 250...
Page 4: ...WD 290 15 tray washer disinfector with automatic glass sliding doors...
Page 5: ...Accessories instrument racks...
Page 8: ...4 key factors for cleaning efficacy Chemicals...
Page 19: ...TOSI testing...
Page 27: ...Dynamic fill advantages...
Page 28: ...WD 290 high voltage section Power Supply Circuit breakers Contactors...
Page 73: ...Drying system Separation of air and water systems...
Page 82: ...Service mode...
Page 83: ...Analysis...
Page 84: ...Dosing...
Page 85: ...Controls...
Page 86: ...Parameter 1 Analysis 2 Dosing 3 Controls 4 Parameter 5 Configuration 6 Limits...
Page 87: ...Program Structure...
Page 88: ...Configuration...
Page 89: ...Limits...
Page 90: ...WD 250 magnet template...
Page 96: ...Process status display...
Page 97: ...Menu...
Page 98: ...Log on...
Page 99: ...Password entry...
Page 100: ...Data entry example...
Page 101: ...Machine information...
Page 102: ...Error messages...
Page 103: ...Settings...
Page 104: ...Measurements...
Page 105: ...System configuration...
Page 106: ...Parameters...
Page 107: ...Function medium...
Page 108: ...Door position...
Page 109: ...Manual function...
Page 110: ...Magnet positions...
Page 125: ...PM check sheet...
Page 126: ...Wash are disassembly...
Page 127: ...Lower wash arm docking port...
Page 128: ...Upper wash arm...
Page 129: ...Daily screen cleaning...
Page 130: ...Wash Arm...
Page 131: ...Water supply strainers...
Page 132: ...Air HEPA pre filter...
Page 133: ...Drying system...
Page 134: ...Air water check valve...
Page 135: ...Air water check valve...
Page 136: ...Light bulb...