16
Display
Summary of function
unit5
Units of display
Enables the format of the displayed
time to be selected.
12 : 00 : 00
Hours, minutes & seconds
12 : 00
Hours & minutes
30 : 00
Minutes & seconds
30
Seconds
Excludes delay of optional control
outputs which is always shown in
seconds and the grand total which is
always shown in hours.
See section 5.5.10
5Et t
Set time
This is the BA578E Timer's setpoint.
When controlling an external event via
the
optional
control
outputs
the
BA578E will time-down from the set
time to zero or time-up from zero to the
set time.
Note:
Timer will only start if a non zero
value is entered for set time
5et t
.
See section 5.5.11
CYCLE5
Repeat timing cycle
Contains a sub-menu with three
sub-functions,
EnbL
,
CYCL Cnt
and
r5t dELA
.
See section 5.5.12
Cycle function enable
EnbL
Enables or disables the cycles function
without changing the parameters.
See section 5.5.13
Cycle count
CYCL Cnt
Defines the number of times that the
timer cycle is repeated.
See section 5.5.14
Restart delay
r5t dELA
Defines the time delay between timer
cycles.
See section 5.5.15
AC5Et t
Access
5Et t
from display mode
Contains
two
sub-functions,
EnbL
which when activated allows the set
time
5Et t
and restart delay
r5t dela
to be adjusted from the display mode.
The second sub-function
ACCd
defines
a separate access code to protect
access to
5Et t
from the display mode.
See section 5.5.16
Display
Summary of function
uP or dn
Direction of count
Defines whether the Timer times-up
from zero to the set time
5Et t
, or
times-down from
5Et t
to zero.
See section 5.5.17
P-FAiL
Power Failure
Defines how the Timer functions when
power is restored after a power failure.
Contains three alternative options
idle
,
Pau5E
and
continue
.
Idle
Idle
Timer returns in stopped state as if
having completed single timing cycle
displaying Timer value when power
was lost. Timing resumes when reset
followed by start instructions are
received.
Pause
pau5e
Timer
returns
in
paused
state
displaying Timer value when power
was lost. Timing resumes when start
instruction is received.
Continue
continue
Timer will continue without any manual
intervention.
See section 5.5.18
LoC r5Et
Local reset
Contains
two
sub-functions
which
when enabled allow the Timer and the
grand total, which represents total
Timer run-time, to be reset to zero via
the front panel push buttons while the
Timer is in the display mode.
See section 5.5.19
Local total reset
r5Et
.
EnbL
When
on
is selected, Timer is reset to
zero, or
5et t
if timing-down, when the
&
and
*
buttons are operated
simultaneously for more than three
seconds in the display mode.
See section 5.5.20
Local grand total reset
CLr Gtot
When
on
is selected the grand total,
which represents total run-time, may
be reset to zero by operating the
)
and
*
buttons simultaneously for
more than 10 seconds in the display
mode.
See section 5.5.21