Step 14 Reset the grand total to zero
Before completing configuration the run-
time grand total should be reset to zero.
Using the
&
or
*
button select
Clr
.
Gtot
in the configuration menu and press
(
which will cause
Clr
.
no
to be displayed.
Again using the
&
or
*
button select
CLr YE5
and press
(
which will result in a
0000
display with one digit flashing. This is
a request for the instruction to be
confirmed by entering
5urE
using the
&
or
*
buttons to set each digit and the
(
button to move control to the next digit.
Pressing
)
will then reset the run-time
grand total to zero and return the
instrument to the
Clr
.
Gtot
prompt in the
configuration menu.
See 6.17.
Step 15 Define the security code
Defining a security code prevents
unauthorised access to the configuration
menu. Using the
&
or
*
button select
CodE
from the configuration menu and
press
(
which will reveal
0000
with the
first digit flashing. This example requires
the security code to be 1209, using the
&
or
*
button set the flashing digit to 1 and
press
(
to transfer control to the second
digit. When all have been entered press
)
to return to the main configuration
menu.
See 6.18.
Step 16 Return to the display mode
Configuration of the BA317NE is now
complete. Pressing the
)
button will
save the new configuration and return the
Tachometer to the display mode. The
BA317NE will display
dAtA
followed by
5AVE
while the new information is stored
in permanent memory, which will be
protected from unauthorised adjustment
by the security code.
To obtain a stable display it may be
necessary to adjust the two filter
parameters and the level of debounce
during commissioning of the Tachometer.