Required tools, ancillary and safety equipment
Tool
Ancillary equipment
Safety equipment
Acid-resistant tools
Screwdriver
Allen key
Open-ended spanner 8, 17
Box spanner 8
Adjustable/pump spanner
Cleaning fluid
Kitchen roll or cotton
cloths
Acid-resistant cleaning
container
Alcohol for lamp glass
Gloves
Safety goggles
Pers. protective equipment
10.1.2
.
Changing the UV-lamp
The UV-lamp should be changed after 8000 operational hours. With the lamp
replacement check or replace the viewing window.
1.
The system should be safely isolated from the electricity supply. Wait for the UV
lamp to cool down (minimum approx. 2 minutes).
2.
Loosen safety screw.
3.
Remove protective cap with gentle rotating motion
4.
The lamp will pull out and carefully disconnect the socket. Do not let lamp fall.
5.
Install new lamp.
Caution
do not touch the UV lamp with bare hands. If this
happens, clean with an alcohol wipe (or similar).
6.
Assemble in reverse order
7.
Check for tightness and firm mechanical/electrical connections
Locking screw
Before restarting the equipment
please ensure
that all screw connections are secure to
ensure no UVC radiation can reach the skin and eyes.