20
2. AC Voltage Measurement
Range
Resolution
Accuracy
45Hz – 400Hz
400Hz – 1kHz
6.000V
1mV
±(1.0%+3LSD)
±(2%+3 LSD)
60.00V
10mV
600.0V
100mV
1000V
1V
±(1.2%+3LSD)
±(2.5%+3 LSD)
Overload protection:
1000V rms
Input impedance:
Around 10M
Ω
Frequency response:
45Hz – 1kHz
Note:
Frequency (on secondary display) may not be displayed if the measured
voltage is below 20% of the display voltage range.
3. Low Pass Filter
Range
Resolution
Accuracy
6.000V
0.001V
45 to 200Hz ± (2%+40 LSD)
200 to 440Hz ± (6%+40 LSD)
60.00V
0.01V
600.0V
0.1V
1000V
1V
Block ac voltage signals above 1KHz
Overload protection:
1000Vp
4. Frequency Measurement
Range
Resolution
Accuracy
60.00Hz
0.01 Hz
±(0.1%+3 LSD)
600.0Hz
0.1 Hz
6.000kHz
1 Hz
60.00kHz
10 Hz
600.0 kHz
100Hz
6.000MHz
1KHz
±(0.1%+3 LSD)
60.00MHz
10KHz
Overload protection:
1000Vp