BRDoc209
RoboBlot - User Manual [en-GB]
Rev: 2 | CRF: 2665
Bee Robotics Ltd
CONFIDENTIAL
Pg. 40 of 64
Upon closing the pop-up, the instrument will prime the lines of the water and wash pumps with
the cleaning solution, then soak the lines for a few minutes. The progress onscreen will display
thus:
After the soaking stage, the screen will prompt the user for further intervention:
The steps for this are:
Discard any remaining cleaning solution
Fill the wash and water bottles with DI Water, and press OK onscreen.
Close door and press OK onscreen to start the cleaning process.