Doc ID: 76955 Rev. 2.0
Chapter 3: Maintenance
17
3. Remove the white drip tray (red oval) by lifting it slightly and
pulling it out:
4. If necessary, appropriately dispose of the liquids in the drip
tray.
5. Use pre-moistened cleaning wipes with 10% bleach or 70%
ethyl alcohol to wipe the sides, top, and bottom of the drip
tray and the tube receptacles for the Waste Collection
Container (Cat. No. 650000090) and Bead Retrieval Tube
(5 mL LoBind Tube).
6. Use a dry lint-free wiper to dry the components.
7. Orient the white drip tray so that its larger compartment faces
upwards and is positioned towards the interior of the
instrument, and then reinsert the drip tray fully until it stops.
The compartments of the drip tray correspond to the tube sizes
above it.
8. Move the front slider with the correct collection tubes to the
appropriate position to
BEADS
or
WASTE
.
9. Dispose of the used cleaning materials in the biohazard
containers.