Specifications
Software Specifications
Flow-Count (
) Method
The Flow-Count (
) Method uses Flow-Count Fluorospheres and a flow cytometer to determine
the absolute counts with the following formula:
Absolute Count (cells/
μ
L) = (Count of Positively Stained Cells of Interest ÷
Count of Flow-Count Fluorospheres) x Flow-Count Fluorospheres Assayed Concentration
This Single Platform Technology (
) method is automated using the Navios software and requires
all of the following:
•
Flow-Count Fluorospheres must be used.
•
The Assayed concentration of Flow-Count Fluorospheres must be entered in the Absolute Count
Calibration dialog box and selected in Acquisition Manager when running a protocol that
includes a CAL region. See
•
The number of fluorospheres counted must be
1000.
Note
: A total white blood cell count is not necessary for determining an absolute count for the cells
of interest if using Flow-Count Fluorospheres.
The percentages and absolute counts are automatically printed on the
Dual Platform Method
The Dual Platform Method calculates absolute counts by combining results from hematology and
flow cytometry and using the following formula:
Absolute Count (cells/
μ
L) = Total White Blood Cell Count (cells/
μ
L) x %Lymphocytes x
%Positively Stained Cells of Interest ÷ 10
4
Absolute counts are obtained using the Report function of the Navios Software. For information on
generating a report, refer to
or the Navios System Help. The percentage and
absolute count are automatically printed on the
Setup Mode
During Setup mode the workspace is a rolling display that is updated up to three times per second.
The maximum number of events can be set on the
Cytometer Control Acquisition Setup Tab
incoming data is not saved.
Acquisition
During data acquisition, the plots are updated in real time. When one plot is displayed with statistics
underneath, the statistics are also updated in real time. Unlimited one or two parameter plots are
available for any given sample.
One-parameter plots have 1,024-channel resolution.
Two-parameter plots have up to 512- x 512-channel resolution.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......