Using Navios Software
Analysis Menu
Override runtime CAL Factor...
Check this option if you want to replay a batch of listmode files using a different Cal Factor.
Advanced
If your Administrator has assigned you access rights, the
Advanced
button is active and you can
enter and modify batch and count information. Enter the
Product name
,
Expiry Date
,
Lot Number/ID
and
CAL Factor.
Add Absolute Calibration Batches
in the User Administration section.
ADD
Use the ADD button to enter new lot information (only available on the Advanced screen).
Replace
Use the Replace button to replace the selected information with updated values (only available on
the Advanced screen).
Delete
Use the Delete button to remove a selected lot from the list (only available on the Advanced screen).
Setting a CAL Region
Create a Region around the absolute count calibration particles, the region MUST be named CAL. Set
or select the calibration factor for the batch of beads in use from the
Analysis >
NOTE
If the Cal Factor is set to 0 or there is no CAL Region present in the protocol, “Error” appears in the
Cells/uL field. If the number of events in the CAL region equal to or less than 1000, "Invalid" appears in
the Cells/uL field.
Calculate Results
This option calculates the statistics for all files selected on the screen.
Refresh
Selecting
View > Refresh
, or pressing
(
F5
)
, causes Navios software to refresh the screen, all plots
and FlowPAGEs. This option is useful when Windows fails to refresh the screen correctly leaving
some graphical errors visible.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......