PN 4277248DD
4-9
SPECIFICATIONS/CHARACTERISTICS
PERFORMANCE CHARACTERISTICS
4
.
NRBC Paired Sample Imprecision
Table 4.19 shows the NRBC results for whole blood specimens with and without NRBC
collected in a salt of EDTA. Each specimen was analyzed in duplicate using Automatic Mode.
The individual differences for each specimen were calculated (run1 - run 2) and the Mean
and Standard Deviation of the Differences were determined.
MCV Accuracy Characteristics
Table 4.20 shows MCV accuracy at temperature extremes.
Table 4.17 Differential Suspect Flagging at Room Temperature
Sample #
1 hour
4 hour
8 hour
16 hour
24 hour
1
Imm Ne 1
Imm Ne 1
Imm Ne 1
Imm Ne 1
Mo Blast, Imm Ne 1,
Imm Ne 2
2
*
*
*
*
Imm Ne 2
3
*
*
*
*
*
4
*
*
*
*
*
5
*
*
*
*
*
*Indicates that there were no flags present.
Table 4.18 Differential Suspect Flagging at Cold Temperature
Sample #
4 hour
8 hour
16 hour
24 hour
36 hour
48 hour
1
Imm Ne 1
*
Imm Ne 1
Imm Ne 1
Imm Ne 1
Imm Ne 1
2
*
*
*
*
*
*
3
*
*
*
PLT Clumps
*
*
4
*
*
*
*
*
*
5
*
*
*
*
*
*
*Indicates that there were no flags present.
Table 4.19 NRBC Imprecision Characteristics,Paired Sample Analysis - Replicate 1 vs. Replicate 2
Parameter
Units
N
Population
Minimum
Population
Maximum
Mean
Difference
SD
of
Difference
NRBC
%
488
0.00
184.44
-0.14
1.69
Table 4.20 MCV Performance Characteristics
Temperature
With LH Series
Reagents
With ISOTON 4
LYSE S 4
@ 60 degrees F
-1.9%
-0.7%
@ 90 degress F
-0.1%
2.4%
Summary of Contents for COULTER LH 750 System
Page 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Page 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Page 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Page 118: ......