OPERATION PRINCIPLES
COUNTING AND SIZING
3
Figure 3.5 Sweep Flow
RBC Size Distribution
After editing, the RED/WHITE PROCESSOR converts the three sets of both the RBC and
WBC pulses. It converts each pulse to a number that corresponds to the size of the cell. The
Analyzer Microcontroller card (AMC) reads this data and generates histograms. The digital
information from each aperture is stored according to volume in 256-channel,
size-distribution histograms.
To ensure that the size-distribution curve accurately reflects the true cell population, the
system extends RBC sensing for up to eight additional 2-second sensing periods whenever the
RBC data accumulations are below a predetermined value. The RBC size distribution curve
reflects the total data accumulated in all of the sensing periods.
After the sensing periods are completed, the system sends these histograms through the
Communication Interface (COMM INTFC) card to the Workstation for display.
Platelet Method
Pulses representing cells from 2 to 20 fL are classified as platelets. The analyzer sorts the
pulses into three 64-channel size distribution histograms. Data collection occurs between a
two second minimum, up to 20 second maximum time, in one second increments. The data
collection stops when at least 1500 cells are collected in each of the three histograms, or 20
seconds is reached.
The analyzer sends this information is sent to the Workstation. At the Workstation, all
histograms are smoothed. The mode channels, and left and right valleys, are identified for
each of the three histograms. Information is analyzed to determine whether normal curve
fitting and counting occurs, and whether special conditions for count correction need to be
considered. For normal curve fitting, the following conditions have to be met for each
histogram:
1.
The PLT is > 20,000 in the raw histogram.
2.
The curve is positive and unimodal.
CELL
SENSING
ZONE
SWIRLING
EFFECT
NO
SWEEP FLOW
A
B
TO
WASTE
SENSING
ZONE
SWEEP
FLOW
CELL
Summary of Contents for COULTER LH 750 System
Page 26: ...PN 4277248DD 1 12 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 54: ...PN 4277248DD 3 22 OPERATION PRINCIPLES XB ANALYSIS ...
Page 84: ...PN 4277248DD 4 30 SPECIFICATIONS CHARACTERISTICS BAR CODE LABEL SPECIFICATIONS ...
Page 90: ...PN 4277248DD 5 6 HAZARDS DISPOSAL OF ELECTRICAL INSTRUMENTATION ...
Page 118: ......