PN 4277248DD
2-5
INSTALLATION
INTERUNIT CONNECTIONS
2
Figure 2.4 Rear of instrument with DB 9-Pin Connector
Note
: The DB 9-pin connector on the computer base, designated as COM 1, is located beside
the video display connector.
Pneumatic/Hydraulic Tubing Connections
shows the updated tubing connections between the Diluter and the:
r
Reagent containers
r
Waste container
r
Power Supply pressure and vacuum supplies.
CAUTION
Possible reagent siphoning effect and priming problems can occur if a reagent container is
placed above the level of the Analyzer. Do not place reagent containers above the level of the Analyzer.
IMPORTANT
Placing Reagent Paks in any location other than on the counter next to the instrument may
cause erroneous results.