PN B07493AD
A-2
Installation
Electrical Requirements
4
Ambient temperatures during operation should not be lower than 10°C (50°F) or higher
than 30°C (86°F).
•
Relative humidity should not exceed 75% (noncondensing).
Figure A.1
Centrifuge Dimensions
Electrical Requirements
To reduce the risk of electrical shock, this centrifuge comes with a 2.5-m (8-ft) three-wire electrical
cord (attached to the power connector at the rear of the instrument) and plug to connect the
centrifuge to earth-ground. The plug for use with 120-VAC, 60-Hz centrifuges is the NEMA 5-15P;
a NEMA 5-15R socket should be available.
cm
in
96.5
38.0
76.2
30.0
7.6
3.0
7.6
3.0
34.3
13.5
62.2
24.5
120-V, 60-Hz centrifuge
108–132 VAC, 12 A, 60 Hz