PN B07493AD
2-6
Operation
Manual Operation
Entering Run Temperature
Run temperature can be set from 2 to +40°C. If no value is entered, the centrifuge selects the last
entered temperature.
1
Press
TEMP
.
•
The cursor and the temperature range (2 to 40°C) appear.
2
Use the keypad or arrow keys to enter required temperature.
•
Entered temperature appears on the display.
•
(If the entered temperature is outside the valid temperature range, an error message
will appear and the input will be cleared.
a.
Enter a valid temperature.)
3
Press the next run-parameter key or
ENTER/SAVE
.
Entering Acceleration and Deceleration Rates
You can select one of ten acceleration rates and eleven deceleration rates to protect the gradient
and sample-to-gradient interface. Select rates depending on the type of run you are performing—
for pelleting runs, where sample mixing is not a concern, maximum acceleration and deceleration
can be used. If running delicate gradients, a lower setting may be needed. Typical acceleration and
deceleration rates are listed in
Table 2.1
. If no rate is selected, the centrifuge automatically
accelerates and decelerates at maximum.
Table 2.1
Acceleration and Deceleration Rates (in Minutes:Seconds) to and from Maximum Speed
a
Rate
SX4750
SX4750
SX4750A
SX4750a
FX6100
(4300 RPM)
(4000 RPM)
(4300 RPM)
4000 RPM
(10,200 RPM)
Accel
Decel
Accel
Decel
Accel
Decel
Accel
Decel
Accel
Decel
10 (MAX)
9
8
7
1:24
2:03
2:14
2:27
1:21
2:07
2:18
2:33
1:14
1:53
2:04
2:16
1:14
1:59
2:08
2:21
1:23
2:00
2:14
2:27
1:21
2:06
2:19
2:30
1:14
1:50
2:04
2:17
1:16
1:58
2:09
2:23
1:24
2:23
3:24
4:28
1:23
2:26
3:25
4:23