PN B07493AD
3-6
Troubleshooting
Retrieving Your Sample in Case of Power Failure
2
Make sure that the rotor is not spinning.
•
No sound or vibration comes from the centrifuge.
WARNING
Never attempt to override the door interlock system while the rotor is spinning.
3
Locate the two joined release cords underneath the bottom center of the centrifuge (not visible
from the centrifuge front). See
Figure 3.1
.
4
Pull on the cords or insert a screwdriver through the two cords and pull to release.
•
The latch releases and the door opens.
a.
If the rotor is still spinning,
close the door and wait until it stops before attempting to remove it
.
WARNING
Never attempt to slow or stop the rotor by hand.
Figure 3.1
Door Latch Override
1.
Release cord
1