General Data Logging Settings
Trigger
Available options:
Peri-
odic
,
Keyboard
and
Clock Start
.
When
Keyboard
is selected, the
fields
Interval
,
Samples
and
Time
are not needed and thus not shown.
When
Clock Start
is selected, a
separate field for entering the start
time is shown. Enter the time in 24
hour format.
Save Method
Op-
tions:
Instant
,
Average
,
Minimum
,
Maximum
,
Min & Max
and
All val-
ues
.
Interval
Accepted values: 1 to 3600 seconds. Enter the value as an integer.
Samples
Maximum amount of samples: 70000.
Time
Automatically calculated based on the
Interval
and
Samples
set-
tings. Alternatively, by changing the
Time
settings, the
Samples
value is automatically changed to suit the new
Time
settings.
Note.
In addition to the previously mentioned limits, the maximum values
of
Samples
and
Time
settings also depend on the available free
memory. If you enter values that require more memory than is avail-
able, MC5P automatically replaces
the entered value with the currently
allowed maximum value.
When a data log is done, you can
enter a name for the results.
Starting the Data Log
To start logging, select
B
/
Data Log-
ging
in the Data Logging configu-
ration window. Start the Logging by
selecting
C
/
Start Logging
in the
Logging window.
10.11.2003 8:14
DATA LOGGING SETUP
1
Trigger
2
ET:RTD(mea) °C Pt100
α
385
3
ET:TCi(mea) °C K NiCr/niAl
4
None
5
None
6
None
7
None
Save Method
Interval [s]
Samples
Time [h min s]
Periodic
Instant Value
60
720
12
Mode
Basic
Logging
Data
Edit
MENU
P3:B bar abs
0
0
Abs
bar
1.0113
10.11.2003 13:12
DATA LOGGING
Pt100
α
385
°C
24.17
K NiCr/NiAl
°C
24.39
Posit. kantti
kHz
-------
Posit. kantti
kHz
-------
Posit. kantti
kHz
-------
Posit. kantti
kHz
-------
1
2
3
4
5
6
7
Count
647
38802 s
Logging
Stop
MENU
Back
Advanced Utilities