115
Calibration
1. Move to Calibration Mode and select the
instrument to be calibrated. If appli-
cable, activate the pressure controller
communication as described in chap-
ter
The Instrument Window
on page
109.
2. Start the calibration as presented in
chapter
A Calibration Procedure Us-
ing MC5P
on page 110.
3. Set the pressure manually using, e.g.
a pressure pump. If a pressure control-
ler is connected to MC5P, the pressure
controller generates the pressure based
on MC5P’s requests.
4. In manual calibration: Accept the points
using the
B
/
Accept Readings
Function
Key. The calibration proceeds automati-
cally if a pressure output controller is
available and the
Calibration Mode
is
set to Automatic. Additional Calibration
Method specific data is in chapters
About Automatic Calibration
and
About Manual Calibration
, both on
page 112.
5. Save or reject the results as is de-
scribed in chapter
A Calibration Pro-
cedure Using MC5P
on page 110.
6. Either do another calibration run or re-
turn to the Instrument window.
Hint!
These instructions can be adapted for
pneumatic pressure trans-
mitters
and
converters,
too. Instead of connecting the output sig-
nal to a terminal in the E module, connect the pressure output sig-
nal to a suitable pressure module.
Warning!
Working with pressure instruments can be hazardous. Only
qualified personnel may use pressure instruments and pres-
sure sources.
Next…
MC5P’s Support for Instrument Adjustment
on page 123.
Calibrating an Instrument