130
Calibration
Calibration Result Windows
The results are displayed in table
format and as a graph. There are
also some additional data display-
ing the statistics of the calibration
as well as environmental informa-
tion that was automatically inserted
during the calibration or it was
manually entered after the calibra-
tion. If any calibration notes were
written, they can also be seen on
one of the calibration result pages.
Use the
B
/
Next Page
, to change
from one result page to another.
The table with numeric results may
include more rows in the table than
what can be displayed. Use the
vertical cursor keys to scroll
through all rows.
How to Choose Which Calibration Run is Viewed
When viewing Calibration results,
press
D
/
MENU
,
1
/
Calibration
History
. A list of saved calibrations
is displayed. Select the calibration
run to be viewed by moving the
cursor (inverted text) to the desired
calibration run and pressing either
the
or the
key..
Deleting Calibration Results
When viewing Calibration results, press
D
/
MENU
.
To remove the current calibration run, press
3
/
Remove Results
.
To remove all calibration results (for the current instrument), press
4
/
Remove All Results
.
11.11.2003 13:33
Page
Next
MENU
Back
0.40
Output Error [% of span]
+
0
-
0%
50%
100%
Maximum Error
Maximum Hysteresis
Maximum Unlinearity
Span Error
-0.170
0.064
0.044
-0.126
Zero Error
-0.044
11.11.2003 13:34
Back
Page
Next
MENU
25.09.2000 10:52 - As Left - Passed
Input
[°C]
0.00
25.00
50.00
75.00
100.00
75.00
50.00
25.00
0.00
Output
[mA]
3.9930
7.9890
11.9834
15.9777
19.9729
15.9875
11.9925
7.9979
4.0032
Error
[%]
-0.044
-0.069
-0.104
-0.139
-0.170
-0.078
-0.047
-0.013
0.020
1
2