4-2
MEGA3 NFPA Medical Gas Notification System
4107 9016 59.03
4.1.4 Toolbar
Running from left to right across the bottom
of the MAIN screen is the toolbar. The
Normal toolbar is available when the alarm is
operating in normal status and shows:
NOTE
:
The Normal toolbar will automatically
reappear when all faults are resolved.
To switch to the Icon toolbar, touch the
HOME button.
Alarm Button (Red)
Figure 41: Fault Status
Figure 38: Combination Alarm Main Screen
Alarm Status (normal)
Home
Button
Location of Panel
Figure 39: Normal Toolbar
Icons on the toolbar:
1. Home button. This icon button toggles to
the MAIN screen.
2. Alarm button. This icon button toggles to
the ACTIVE ALARM screen.
3. Configuration button. This icon button
toggles to the CONFIGURATION screen.
4. Information button. This icon button
toggles to the INFORMATION screen.
5. History button. This icon button toggles
to the HISTORY screen. This screen is only
available after upgrading to a TotalAlert
Infinity™ series of alarms.
The button that corresponds to the screen
currently in use will be blue.
4.1.5 Fault Status
If a gas sensor is measuring above or below
the alarm level, the high or low pressure in-
dicator on the gas badge will flash for the gas
pressure out of range.
If a gas sensor becomes disconnected or
shorted, both high and low indicators for
that gas will flash and an error code and mes-
sage will be displayed.
If a configured source signal becomes an
open circuit and the source badge and the
signal status will change from green to blink-
ing red.
The Icon toolbar will appear, and the Alarm
button will be flashing red. By default the
screen will change to the Active Alarm
Screen.
Users can touch any flashing red light to go to
the ACTIVE ALARM screen.
Figure 40: Icon Toolbar
Home Button Information Button
History Button
Configuration
Button
Alarm
Button