5
Part No. 6-847706-00 Rev. B00
Wall Outlets
Gemini wall outlets are identified by a Beacon Medical or BeaconMedæs logo on right hand side of
outlet.
Diamond
®
III/DISS III wall/ceiling outlets may be identified by the words “Hill-Rom” or “Medæs” cast
into cover plate on lower right hand side and Model No. 300 series or 400 series on back side of cover
plate.
Diamond
®
II wall/ceiling outlets are identified by Model A30 series on back side of cover plate.
Gemini Console Front Body
Before servicing medical gas outlets, it is very important to identify exact type and vintage of outlet.
Console Outlets
Remove console cover plate to expose front body.
Diamond
®
III Console Front Body
Gemini console outlets will appear as shown
above. Front body will have a solid rectangle of
color coded plastic approximately 1 7/8” wide x
2 3/4” tall. Raised circular keying portion of front
body will rise approximately .200”.
Diamond
®
III console outlets will appear as
shown above. Front body will be easily identified
by appearnace of galvanized front body plate
visible in corners.
Outlet Identification
Color Coded
Plastic
Galvanized
Front Body
Plate
Upgrade of Diamond II
®
Medical Gas Outlets
Rough-in assembly must be upgraded to accept Gemini front body. This requires installation of a new
spring, secondary check, o-ring, rubber seal and retaining ring. Order one Secondary Check Repair Kit,
P/N 6-290802-PG for pressure service or 6-290803-PG for vacuum or WAGD service, and 10 each of
Secondary Check, P/N 6-525344-00 for every 10 outlets to be upgraded Order one each of Gemini or
DISS finish assembly (ref. page 6) for each corresponding gas service.
Diamond
®
II medical gas outlets must be upgraded to Gemini Medical gas outlets.
Diamond
®
II Medical Gas Outlets