7

Part No. 6-847706-00 Rev. B00

Lock ring

Service

Notify appropriate hospital personnel when

performing service or maintenance to medical gas
systems.  Consult Troubleshooting Guide to
determine required repair.  Outlets incorporate a
primary valve and secondary check.  Secondary
check prevents system gas from escaping when
primary valve is removed for service.  Refer to
Replacement Components for identification of
replacement part numbers.

Quick-Connect and DISS Valve Assemblies

(Primary Valve)

Quick-connect and DISS valves may be

removed for replacement of seals and some
components without interrupting service to other
outlets in piping system.  To check a quick connect
or DISS valve for leaks, apply leak detector
solution at primary valve opening both with and
without adapter connected.  Bubbles indicate
pressure leaks. Large leaks may be audible.

Front Body

Diamond

®

III/DISS III outlets can be

disassembled for servicing and replacement
of parts as illustrated in exploded parts
breakdown (pages 10 -12). 

Adapter locking spring on Diamond

®

III front

body housing, and front body o-ring, can be
replaced by dismantling housing as follows: 
1. Loosen two front body screws.

2. Carefully remove front body assembly from

back body assembly.

3. Using needle-nose pliers, crush inside

flange of brass eyelets and remove from
assembly.

4. Remove and replace all worn or damaged

parts. 

5. New brass eyelets must be used when re-

assembling front body assembly. Insert
eyelets and flare rear of each with 60°
tapered punch. 

Back Body

Remove internal parts from back body assembly
by lifting inside ( 1" ) lock ring from rear housing
with small screwdriver, as illustrated below.
Carefully work rubber bushing from back body with
screwdriver. Remove check unit and spring.
Replace components as necessary and
reassemble outlet.

NOTE:

On recessed and surface mount outlet models,
screws are held in place by rubber washers;
console models use elastic stop nuts. 

WARNING:

Be certain gas supply to outlet is shut off
before disassembling back body.

NOTE:

Gas supply need not be turned off when
removing front body - secondary check and
bushing in back body prevent flow from outlet. 

NOTE:

Replace all elastomer seals whenever outlets
are disassembled.

Summary of Contents for Gemini Diamond II

Page 1: ...Service Manual Gemini Diamond II and Diamond III Medical Gas Outlets for Concealed Piping and Console Applications MAN 01 049 Andersen Medical Gas 12 Place Lafitte Madisonville LA 70447 http www TheMe...

Page 2: ...als 4 Cleaning and Lubrication 4 Inspection and Testing 4 Outlet Identification 5 Diamond II Medical Gas Outlets 5 Outlet Assembly Ordering 6 Service 7 Troubleshooting Guide 8 Replacement Components 9...

Page 3: ...t Description DISS outlets are typically used in ceilings of operating rooms or in place of quick connect outlets Connection to outlet is by means of a nut and nipple Gas specific indexing to prevent...

Page 4: ...tient room surfaces that is compatible with acetal and nylon plastic anodized aluminum and die cast zinc Lubricate elastomer seals sparingly with Krytox GPL205 Refer to Tools and Service Materials DO...

Page 5: ...color coded plastic approximately 1 7 8 wide x 2 3 4 tall Raised circular keying portion of front body will rise approximately 200 Diamond III console outlets will appear as shown above Front body wil...

Page 6: ...rbon Dioxide 6 234520 00 6 234720 00 Gemini Oxygen 6 234510 00 6 234610 00 Concealed Nitrous Oxide 6 234511 00 6 234611 00 Quick Connect Air 6 234512 00 6 234612 00 Vacuum 6 234513 00 6 234613 00 WAGD...

Page 7: ...2 Adapter locking spring on Diamond III front body housing and front body o ring can be replaced by dismantling housing as follows 1 Loosen two front body screws 2 Carefully remove front body assembly...

Page 8: ...e DISS nipple Replace O Ring seal Symptom Possible Cause Corrective Action Adapter will not engage and or release Damaged adapter Wrong adapter Damaged frontbody assembly Replace adapter Check adapter...

Page 9: ...ing Ring pkg of 10 Concealed Wall Outlet Back Body 1 2 4 3 5 Console Wall Outlet Back Body 1 2 4 3 5 Repair Kits Below Repair Kits Contain Components to Repair 10 Backbodies Part Number Description 6...

Page 10: ...ch of 6 290804 PG and 1 each of 6 831344 PG shown as item 6 Diamond III To repair 10 outlets order 1 each of 6 290804 PG and 1 each of 0203 5643 870 shown as item 6 Part Number Description 6 290804 PG...

Page 11: ...pkg of 10 2 0221 1587 870 Piston pkg of 10 3 6 622589 PG O Ring Red pkg of 10 4 6 415038 PG Washer pkg of 10 5 6 833953 PG Lock in Adapter Spring pkg of 10 6 0203 5643 870 Eyelet pkg of 50 7 0140 632...

Page 12: ...cuation and WAGD pkg of 10 6 622586 PG For N2O Air N2 and CO2 pkg of 10 Refer to page 6 for complete replacement assemblies DISS Concealed Wall Outlet 1 2 DISS Console Outlet 1 2 Repair Kits To repair...