3
Part No. 6-847706-00 Rev. B00
BeaconMedæs medical gas and vacuum quick-
connect outlets provide a means of connecting to
medical gas piping system. When appropriate
adapter is inserted into outlet, primary valve is
opened, permitting gas flow. Mechanical latch
secures adapter in outlet. Primary seal closes
when adapter is removed.
Outlets are typically for oxygen, nitrous oxide,
air, or vacuum and have been cleaned for
indicated service.
General Product Description
DISS outlets are typically used in ceilings of
operating rooms or in place of quick-connect
outlets. Connection to outlet is by means of a nut
and nipple. Gas-specific indexing to prevent
accidental connection to wrong service is
accomplished by a series of interference
diameters in accordance with Diameter-Index
Safety System (DISS) published by Compressed
Gas Association (CGA) in Standard V-5. Nipple
opens primary valve as nut is tightened to secure
nipple in outlet. Primary valve closes when nipple
is removed.
Gemini Concealed Quick-Connect Front Body
Gemini Concealed DISS Front Body
WARNING:
Do not use oil on adapters or outlets.