BDDF00001 Issue 5
38/56
Alaris™ Syringe Pumps
Specifications
VTBI Complete Rate
Stop, KVO (0 .1ml/h to 2 .5ml/h), set rate if lower than KVO or continue at set rate .
Near End Of Infusion Alarm
1min - 15min to end of infusion, or 10% of syringe volume, whichever is smaller .
End Of Infusion (EOI) Alarm
0 .1% - 5% of syringe volume
Maximum Pumping Pressure Limit
Highest alarm level 1000mmHg (nominal at L-10)
Occlusion Accuracy without pressure set (% of full scale)*
Pressure mmHg
L-0
approx . 50mmHg
L-3
approx . 300mmHg
L-5
approx . 500mmHg
L-10
approx . 1000mmHg
Temp . 23°C
±18%
±21%
±23%
±28%
Occlusion Accuracy with pressure set (% of full scale)*
Pressure mmHg
0
25
500
1000
Temp . 23°C
±2%
±4%
±5%
±6%
Temp . 5°C-40°C
±4%
±7%
±7%
±10%
* - Using most common 50ml syringes under normal conditions (95% confidence / 95% of pumps) .
System Accuracy
Rate
Typical
Pump Specification
≥ 1ml/h
± 2%
± 2%
< 1ml/h
± 2%
± 10%
• Derating - Tempe/- 0 .5% (5 - 40ºC), High Rates +/-2 .0% (rates > syringe volume/h eg . >50ml/h in a 50ml syringe .)
w
System accuracy is +/-2% typical by volume as measured using the trumpet curve test method defined in
EN/IEC60601-2-24 at rates of 1.0ml/h (23ºC) and above when the Pump is used with the recommended syringes.
Caution: Infusion volume accuracy may be compromised at rates below 1.0ml/h. Differences in factors such as size and
plunger force in recognised syringes can cause variations in accuracy and trumpet curves. See also 'trumpet curves'
section in this manual.
Electrical Classification
Class I product . Continuous Mode Operation, Transportable
Data set Specification
A maximum of 30 profiles can be set with a maximum of 100 drugs per profile, with only one profile permitted to be uploaded to a
Non-Guardrails Pump at a time . See Alaris Editor Software DFU for more details .
Battery Specifications
Rechargeable sealed NiMH . Automatically charges when the Pump is connected to AC power .
Mean Time To Power Down from fully charged @ 5ml/h and 23°C ± 2°C under normal conditions is 6 hours*
*95% lower confidence interval of 5 hours 50 minutes
Charging takes 2½ hours from discharge to 90% charge .
Memory Retention
The electronic memory of the Pump will be retained for at least 6 months when not powered up .
Fuse Type
2 x T1 .25L250V
AC Power Supply
115 - 230VAC, 50 - 60Hz, 30VA (under maximum charging conditions) 10VA (nominal) .