BDDF00001 Issue 5
32/56
Alaris™ Syringe Pumps
Configured Options
Configured Options
This section comprises of a list of options which are configurable . Some can be entered via the Pump configurations (available in
Technician Mode) and others through the Alaris Editor Software .
w
Access codes should only be entered by Qualified Service Personnel.
Use Alaris Editor to configure general options, drug library and units enabled for each profile and to configure Syringe Brands and
Models to be enabled .
Alarm Presets
Pumps with version 4 .3 .x software have 2 alarm tones to choose from during configuration:
•
ORIGINAL ALARMS
: Low, medium and high priority alarm tones that sound like the auditory alarms and warnings from software
versions prior to 4 .3 .x
•
3RD EDITION ALARMS
: Low, medium and high priority alarm tones in accordance with IEC 60601-1-8: 2012 and IEC 60601-2-
24:2012
Enter the access code on the Pump for Alarm Presets, see the
Technical Service Manual
or
Information Notice
for details .
1 . Use the
f
keys to select alternative alarm tones .
2 . When the desired alarm tone has been selected press
OK
softkey .
3 . When all modifications have been carried out press
QUIT
softkey .
w
All pumps in a single care area should be configured with the same alarm tones to avoid User confusion.
The Hospital/Facility is responsible for selecting and configuring the desired alarm scheme.
Alaris Gateway Workstation (Workstation) with software versions 1.1.3, 1.1.3 MR, 1.1.5, 1.2, 1.3.0, 1.6.0 or 1.5 do not
support the new Pump low priority visual alarms scheme defined in IEC 60601-1-8: 2012. For Pumps with version 4.3.x
software or higher docked into these Workstations there will be a mismatch of alarm priority displayed. As a result,
Near End Of Infusion, AC Power Fail, Add Drug Not Complete, and Attention alarms will display as a medium visual
priority alarms on the Workstation beacon and a low priority alarm on the Pump. Additionally, for certain information
signals, e.g. those associated with Add Drug Not Complete and Titration Not Confirmed, the Workstation beacon will
illuminate while the beacon on the pump will not. In the event of an alarm priority mismatch, the User should refer to
the alarm on the Pump for the correct priority.
Configured Options
Enter the access code on the Pump for Configured Options, see the
Technical Service Manual
for details .
Clock Set
1 . Select
CLOCK SET
from the Configured Options menu using the
f
keys and press the
OK
softkey .
2 . Use the
f
keys to adjust the date displayed, pressing the
NEXT
softkey to access the next field .
3 . When the correct time and date are displayed press the
OK
softkey to return to the Configured Options menu .
Language
This option is used to set the language of messages shown on the Pump display .
1 . Select
LANGUAGE
from the Configured Options menu using the
f
keys and press the
OK
softkey .
2 . Use the
f
keys to select the language .
3 . When the desired language has been selected press
SELECT
softkey to return to the Configured Options menu .
Contrast
This option is used to set the contrast on the Pump display .
1 . Select
CONTRAST
from the Configured Options menu using the
f
keys and press the
OK
softkey .
2 . Use the
f
keys to select a contrast ratio value . The contrast of the display will change when scrolling through the numbers .
3 . When the desired value has been reached press the
OK
softkey to return to the Configured Options menu .