BDDF00001 Issue 5
24/56
Alaris™ Syringe Pumps
Basic Features
Dosing Summary
To review currently selected dosing information:
1 . Press the
d
button to first access the options menu .
2 . Select
DOSING SUMMARY
.
3 . Review the information and then press the
QUIT
softkey .
Set VTBI over Time
This option allows a VTBI and delivery time to be specified . The rate necessary to deliver the required volume within the specified time is
calculated and displayed .
1 . Stop the infusion . Press the
d
button to access the options menu .
2 . Select the
SET VTBI OVER TIME
option using the
f
keys and press the
OK
softkey .
3 . Adjust the volume to be infused using the
f
keys . When the desired volume has been reached press the
OK
softkey .
4 . Enter the time over which the volume is to be infused . The infusion rate will automatically be calculated . Press the
OK
softkey to
enter the value .
5 . Select the rate at VTBI end from the list using the
f
keys and press the
OK
softkey . The default is
STOP
.
24 Hour Log
This option allows the 24 hour log of volume infused to be reviewed .
1 . Press the
d
button to access the options menu .
2 . Select the
24H LOG
option using the
f
keys and press the
OK
softkey .
The display shows the hourly volume infused . The volume infused shown in brackets is the total volume infused since the volume was
last cleared . See example below:
07:48 - 08:00 4 .34ml (4 .34ml)
08:00 - 09:00 2 .10ml (6 .44ml)
09:00 - 10:00 2 .10ml (8 .54ml)
VOLUME CLEARED
3 . Press the
QUIT
softkey to exit the log .
Event Log
This option allows the event log to be reviewed, if enabled .
1 . Press the
d
button to access the options menu .
2 . Select the
EVENT LOG
option using the
f
keys and press the
OK
softkey .
3 . Scroll through the log using the
f
keys . Press the
QUIT
softkey to exit the log .
Note:
When the event log reaches full capacity, the oldest events will be overwritten by the most recent events .
Data Set Details
To review currently selected data set information:
1 . Press the
d
button to access the options menu .
2 . Select
DATA SET DETAILS
.
3 . Review the information and then press the
QUIT
softkey .
Infusion Setup
To change Infusion Setup
1 . Stop the infusion . Press the
d
button to access the options menu .
2 . Select
INFUSION SETUP
.
3 . Select Infusion Setup required and press the
OK
softkey .
Pump Details
To review Pump information .
1 . Press the
d
button to access the options menu .
2 . Select
PUMP DETAILS
.
3 . Review the information and then press the
QUIT
softkey .
Note:
The following information will be displayed:
• UNIT REFERENCE
An identifier configured in Technician Mode by qualified technical personnel .
• SN
The Serial Number of the Pump
• S/W
Software version of the Pump