13
Recommended separation distance between portable and mobile RF communications equipment and REF 5513E/5514 series
REF 5513E/5514 series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled� The customer or the user of
REF 5513E/5514 series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and REF 5513E/5514 series as recommended below, according to the maximum output power of the communications equipment�
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 0.35 √P
800 MHz to 2.7 GHz
d = 0.7 √P
0�01
0�12
0�035
0�07
0�1
0�38
0�11
0�22
1
1�2
0�35
0�7
10
3�8
1�1
2�2
100
12
3�5
7
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer�
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies�
NOTE 2: These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people�
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy� To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered� If the measured field strength in the location in which REF 5513E/5514 series is used exceeds the applicable
RF compliance level above, REF 5513E/5514 series should be observed to verify normal operation� If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating REF 5513E/5514 series�
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m�
Summary of Contents for 5513E
Page 2: ...2 1 2 3 4 5...
Page 136: ...136 BD 3 1 3 40 4 5 30 PHMB BD 4406 4403A 4412A BD 5513E BD...
Page 138: ...138 2002 96 2006 66 10 C 60 C 10 100 106 kPa 50 kPa B II...
Page 139: ...139 5513E 5514 4406 4403A 4412A 5513E BD 1 m 3 3 30 1 60 C 1 6 6 BD bd com...
Page 140: ...140 6 A 1 800 8 BATTERY BD 5513E 5514 2 BD BD BD BD 4406 4403A 4412A BD BD 6...