Änderungen vorbehalten – Stand 08.05.2019
25
NOTICE about treatment with light:
The selective treatment with light is used to support the treatment with microcurrent. From
practical experience, a positive effect of the light treatment can be derived. However, there
are currently no clinical studies available that prove the effectiveness of light treatment.
The light channel runs synchronous to channel a, that means the current signal output by
channel A is output on channel L as light pulses via the light hand piece.
Connection of Patient and Light Cables:
Fig. 5 – Connection of Patient and Light Cables
When connecting a patient cable or light hand piece, it is important to observe the correct
orientation of the connector. The arrow printed on the connector has to face upward and be
aligned with the white mark on the socket in the device. Under no circumstances may the
connector be forced into the socket! The connector snaps into the socket. When removing
the connector, hold it on its sides with two fingers and pull straight back out of the socket. Do
not twist or use excessive force to remove the connector.