Chapter 4: Operating Instructions
4-10
Clinician’s Operation Manual
Choosing the Sensor
WARNING! Use only SpO
2
sensors and cables supplied with, or specifically intended for use
with, this BCI® oximeter. Use of sensors not intended for use with this device
may cause inaccurate readings.
WARNING! Incorrectly applied sensors may give inaccurate readings.
f
Refer to the
sensor insert for proper application instructions.
Choose the appropriate sensor from the following chart. Select the sensor based on the patient’s
size, available application site, attachment method and other pertinent clinical information.
f
See the sensor’s instruction insert for detailed attachment methods and other important
information.
BCI® Sensors
PATIENT
SITE
DESCRIPTION
Adult
over 45 kg
Finger
Finger or Toe
Ear
3044: Sensor, Reusable, Adult
3044S: Spot Check Sensor, Reusable, Adult
3444: Sensor, Reusable, Comfort Clip
®
3043: Sensor, Reusable, Universal ‘Y’
1300: Sensor, Disposable, Adult Finger
7
WW3078: Sensor, Reusable, Ear
Pediatric
15-45 kg
Finger
Finger or Toe
Ear
3044: Sensor, Reusable, Adult (>20 kg)
3044S: Spot Check Sensor, Reusable, Adult
3444: Sensor, Reusable, Comfort Clip
®
3178: Sensor, Reusable, Pediatric Finger (5-45 kg)
3178S: Spot Check Sensor, Reusable, Pediatric
3043: Sensor, Reusable, Universal ‘Y’
1301: Sensor, Disposable, Pediatric Finger
7
WW3078: Sensor, Reusable, Ear
Infant
3-15 kg
Hand or Foot
Toe
Finger or Toe
3043: Sensor, Reusable, Universal ‘Y’
3025: Sensor, Reusable, Wrap, Infant
1303: Sensor, Disposable, Infant
7
All
Extension Cable
3311: Cable, Oximetry 1.5 meters (5 feet)
NEONATE SENSORS FOR THE WW1020 ONLY
Neonate
under 3 kg
Hand or Foot
Foot
1302: Sensor, Disposable, Neonate/Small Infant
7
3026: Sensor, Reusable, Wrap, Neonate/Small Infant
SMALL INFANT SENSORS FOR THE WW1000 ONLY
Small Infant *
under 3 kg
Hand or Foot
Foot
1302: Sensor, Neonate/Small Infant (disposable)
7
3026: Sensor, Wrap, Neonate/Small Infant (reusable)
*The BCI® 1302 and 3026 oximetry sensors should not be used on neonatal patients with the
WW1000 monitor. Testing has not been conducted for the WW1000 for patients less than 30 days
old. These sensors may, however be used on older patients.