Chapter 11: Specifications
11-6
Clinician’s Operation Manual
Pulse Amplitude Index for the WW1020 Oximeter
Range:
0.03% to 20.00%
6
(0.01% increments)
Pulse Amplitude Index is defined as PI = (100 * AC)/DC where AC is
the alternating current (pulsatile component of the signal) and DC is
direct current (non-pulsatile component of the signal). It is a relative
measure of pulse-signal strength over time at a pulse oximeter
monitoring site, and is non-pulsatile in nature.
Display:
9-segment green / yellow LED bargraph, logarithmic scale
Display Update Rate:
10 Hz
The oximeter will display dashes (Invalid data) for SpO
2
and PR when the PI value is below 0.03%.
Low perfusion performance is verified by simulator testing.
6
As tested with Industry Standard Simulator.
The PI value maps to a 9-segment bar graph as shown below. The two lowest bars (1&2) are bi-
color (Yellow & Green), and bars 3 through 9 are Green.
PI From
Oximeter
0.00-
0.07
0.08-
0.15
0.16-
0.31
0.32-
0.63
0.64-
1.27
1.28-
2.55
2.56-
5.11
5.12-
10.23
10.24-
20.47
Segments
Lit
1Y
2Y
3G
4G
5G
6G
7G
8G
9G
Pulse Amplitude Index for the WW1000 Oximeter
Range:
0.00% to 20.00%
5
(0.2% increments)
Pulse Amplitude Index is defined as PI = (100 * AC)/DC where AC
is the alternating current (pulsatile component of the signal) and
DC is direct current (non-pulsatile component of the signal). It is
a relative measure of pulse-signal strength over time at a pulse
oximeter monitoring site, and is non-pulsatile in nature.
Display:
9-segment green / yellow LED bargraph, logarithmic scale
Display Update Rate:
10 Hz
5
As tested with Industry Standard Simulator.
The PI value maps to a 9-segment bar graph as shown below. The two lowest bars (1&2) are bi-
color (Yellow & Green), and bars 3 through 9 are Green.
PI From
Oximeter
0.00
>0 - 0.2
>0.2
- 0.4
>0.4
- 0.6
>0.6
- 1.2
>1.2
- 2.6
>2.6
- 5.0
>5.0
- 10.2
>10.2
Segments
Lit
1Y
2Y
3G
4G
5G
6G
7G
8G
9G
Printer
Paper:
58mm (2.28 inches), Thermal
Modes:
Real Time, Tabular Trend, Graphic Trend
Paper Speed:
12.5 mm/sec
Trend Graphs:
Headers print with graphs