BC Biomedical FINGERSIM Instruction And Service Manual Download Page 5

 

IV. THEORY 

OF 

OPERATION 

 

The principle of differential light absorption is used by a pulse oximeter to determine the oxygen saturation of 
arterial blood (SpO

2

).  Red light and infrared light are differentially absorbed by oxygenated and deoxygenated 

hemoglobin.  The pulse oximeter has a sensor with light emitting diodes (LEDs) that provides these wavelengths of 
light for transmittance through a measurement site, usually a finger.  Based on the relative absorption of these two 
wavelengths of light at the measurement site, the pulse oximeter determines the relative amount of oxygenated and 
deoxygenated hemoglobin, which is calculated as SpO

2

.  

 

In order to make this calculation independent of skin color, finger size, etc., the pulse oximeter uses only the time 
varying light absorption component generated by the patient’s pulse.  In addition, the pulse oximeter uses the period 
of pulsation to measure the pulse rate.  

 

The FingerSim™ absorbs light very much like a human finger.  The overall red and infrared light absorption of the 
FingerSim™ approximates the overall light absorption of a typical finger.  In addition, the red and infrared photo 
spectrometric light absorption of the inner solution approximates arterial blood as seen by the oximeter at 80%, 90% 
and 97% oxygen saturation levels (see Figures 3, 4, and 5).   

 

Minor SpO

2

 variations will be seen between oximeter manufacturers because standards correlating red and infrared 

light absorption to oxygen saturation in pulse oximetry are not available.  Each manufacturer has developed its own 
correlation and inevitably some differences have developed (see “Health Devices” June 1989).  In addition, minor 
SpO

2

 variations between sensors will be observed due to the fact that red and infrared emitting light sources vary 

slightly between sensors.   

 

The FingerSim™ enables the healthcare professional to repeatedly test and evaluate the pulse oximeter system 
(oximeter and sensor) under controlled light absorption conditions. 

 

The time varying light absorption component required by a pulse oximeter is created in the FingerSim™ by 
rhythmically pressing the color coded end.  This creates a volume change in the distal (sensor) end of the 
FingerSim™, analogous to the heart creating blood pressure waves that force blood into the finger.  The amplitude 
and rate of the pulse wave can be varied by changing the applied pressure and interval. 

Summary of Contents for FINGERSIM

Page 1: ...M PULSE OXIMETER TEST SYSTEM BC Biomedical BC Group International Inc 9415 Gentry Ave St Louis MO 63125 PROBLEMS Check Section XII Trouble Shooting Chart then if still troubled call BC Biomedical Cust...

Page 2: ...ICATIONS 11 A FingerSim 11 B Holder 11 IX SERVICE AND MAINTENANCE 12 X WARRANTY 12 XI ACCESSORIES 12 XII TROUBLE SHOOTING CHART 13 FIGURES Figure 1 FingerSim 4 Figure 2 Replacement Holder 4 Figure 3 8...

Page 3: ...possible operating conditions a pulse oximeter may encounter Use the FingerSim as an adjunct to other indications to determine proper pulse oximeter operation CAUTION Improper insertion of the Finger...

Page 4: ...od as measured by an oximeter at different oxygen saturation values nominally 97 90 and 80 The 97 90 and 80 FingerSims are easily identified by the color coded end caps Red 97 Blue 90 and Black 80 Squ...

Page 5: ...typical finger In addition the red and infrared photo spectrometric light absorption of the inner solution approximates arterial blood as seen by the oximeter at 80 90 and 97 oxygen saturation levels...

Page 6: ...RB ANCE Figure 3 80 SpO2 Absorption Spectrum Figure 4 90 SpO2 Absorption Spectrum Figure 5 97 SpO2 Absorption Spectrum FingerSim LIQUID 90 SpO2 NOMINAL 0 4 620 720 820 920 WAVELENGTH nm ABSORB ANCE Fi...

Page 7: ...SpO2 value For example if the 80 FingerSim was being used at 90 F ambient temperature then the expected simulation would be increased to 83 74 75 76 77 78 79 80 81 82 83 84 85 86 40 45 50 55 60 65 70...

Page 8: ...lder see Figure 9 Figure 9 Sensor Cable Routing Through Holder C Sensor Alignment and Position The light absorption characteristics of the FingerSim are specified for light passing directly through th...

Page 9: ...Pulse Rate readings WARNING FingerSim are fragile and must be handled with care they contain glass Figure 10 Positioning the FingerSim in the Holder B Attach the Sensor Attach the sensor under test to...

Page 10: ...ngs between manufacturers is possible due to each manufacturer s interpretation of how the SpO2 value relates to the absorption of the red and infrared light in blood see Theory of Operation Figure 12...

Page 11: ...c 20dB to 40dB Operating Temperature Range 65 F to 90 F Long Term Storage Temperature Range 32 F to 104 F Typical infrared percent modulation when squeezed 0 to 5 Red to Infrared Ratio a c 72 5 F and...

Page 12: ...the applicable warranty period i e before the calibration date This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any FingerSims or accessories delivered to the purch...

Page 13: ...or leaking FingerSim Replace FingerSim Outside calibration date Replace FingerSim Improper sensor attachment to the FingerSim Ensure the sensor is attached to the FingerSim per section VIII B Damaged...

Reviews: