XII. TROUBLE SHOOTING CHART
Before calling Customer Support, please check the following chart for a possible solution to the problem you are
experiencing.
SYMPTOMS
POSSIBLE
CAUSE
POSSIBLE
SOLUTION
• Oximeter measures slightly
different SpO
2
values than the
nominal FingerSim™ value.
• Expected variation due to
manufacturing tolerances.
• Verify that minor variations
in SpO
2
are not clinically
significant.
• Oximeter measures high or
low SpO
2
bias compared to
the nominal FingerSim™
value.
• FingerSim™ is not at room
temperature.
• Allow at least 1 hour
stabilization at room
temperature .
• Room temperature not at 72.5° F.
• Adjust expected SpO
2
value
(see Charts 6, 7, and 8).
• Cracked or leaking FingerSim™.
• Replace FingerSim™.
• Outside calibration date.
• Replace FingerSim™.
• Improper sensor attachment to the
FingerSim™.
• Ensure the sensor is
attached to the FingerSim™
per section VIII(B).
• Damaged Sensor
• Try another sensor.
• Damaged Oximeter
• Try another oximeter.
• Oximeter displays erratic
SpO
2
values.
• Motion of the sensor relative to the
FingerSim™.
• Use holder provided. Route
cable per VII(B). Be certain
sensor is seated properly on
FingerSim™.
• Damaged Sensor
• Try another sensor.
• Damaged Oximeter
• Try another oximeter.
• No pulse when the
FingerSim™ is squeezed.
• Cracked or leaking FingerSim™.
• Replace FingerSim™.
• Damaged Sensor
• Try another sensor.
• Damaged Oximeter
• Try another oximeter.