BC Biomedical FINGERSIM Instruction And Service Manual Download Page 12

 

12 

IX. SERVICE 

AND 

MAINTENANCE 

 

There is no service or maintenance possible with the FingerSim™ test systems.  Each FingerSim™ is dated to 
identify its useful life.  Replace any FingerSim™ that shows signs of leakage. 

To clean FingerSim™  wipe with isopropyl alcohol. 

 
 

X.

 

WARRANTY 

 

BC GROUP INTERNATIONAL, INC. (BC Biomedical) warrants to the purchaser that each FingerSim™ shall be 
free of defects in materials and workmanship such that each system when properly used shall perform to 
specifications supplied within this manual until the calibration date marked on the FingerSim™ has elapsed.  Note:  
This warranty specifically excludes any internal glass breakage.  BC Biomedical shall replace all FingerSims™ or 
accessories found to be defective in accordance with this warranty, free of charge, for which BC Biomedical has 
been notified by the purchaser that there is a defect, provided said notification occurs within the applicable warranty 
period (i.e. before the calibration date).  This warranty shall be the sole and exclusive remedy by the purchaser 
hereunder for any FingerSims™ or accessories delivered to the purchaser which are found to be defective in any 
manner whether such remedies be in contract, tort or by law. 

The FingerSim™ contains a substance prepared to very exact proportion and are assembled under precise 
specifications.  They are not repairable and must be replaced at the first sign of deterioration or damage.  Any sign 
of tampering or any kind of misuse or abuse shall void the warranty in its entirety. 

 

DISCLAIMER/EXCLUSIVITY OF WARRANTY.  THE EXPRESS WARRANTIES SET FORTH IN THIS 
MANUAL (SECTION X) ARE EXCLUSIVE AND NO OTHER WARRANTIES OF ANY KIND, 
WHETHER STATUTORY, WRITTEN, ORAL, OR IMPLIED INCLUDING WARRANTIES OF FITNESS 
FOR A PARTICULAR PURPOSE OR MERCHANTABILITY SHALL APPLY. 

 

 

XI. ACCESSORIES 

 

The following accessories function with the FingerSim™ Oximeter Testing System: 

 

 

FingerSim™ Replacement Set 

Includes a set of three FingerSim™, 97%, 90% and 80% SpO

2

 (Nominal). 

Replacement Holder 

FingerSim™ Holder 

Summary of Contents for FINGERSIM

Page 1: ...M PULSE OXIMETER TEST SYSTEM BC Biomedical BC Group International Inc 9415 Gentry Ave St Louis MO 63125 PROBLEMS Check Section XII Trouble Shooting Chart then if still troubled call BC Biomedical Cust...

Page 2: ...ICATIONS 11 A FingerSim 11 B Holder 11 IX SERVICE AND MAINTENANCE 12 X WARRANTY 12 XI ACCESSORIES 12 XII TROUBLE SHOOTING CHART 13 FIGURES Figure 1 FingerSim 4 Figure 2 Replacement Holder 4 Figure 3 8...

Page 3: ...possible operating conditions a pulse oximeter may encounter Use the FingerSim as an adjunct to other indications to determine proper pulse oximeter operation CAUTION Improper insertion of the Finger...

Page 4: ...od as measured by an oximeter at different oxygen saturation values nominally 97 90 and 80 The 97 90 and 80 FingerSims are easily identified by the color coded end caps Red 97 Blue 90 and Black 80 Squ...

Page 5: ...typical finger In addition the red and infrared photo spectrometric light absorption of the inner solution approximates arterial blood as seen by the oximeter at 80 90 and 97 oxygen saturation levels...

Page 6: ...RB ANCE Figure 3 80 SpO2 Absorption Spectrum Figure 4 90 SpO2 Absorption Spectrum Figure 5 97 SpO2 Absorption Spectrum FingerSim LIQUID 90 SpO2 NOMINAL 0 4 620 720 820 920 WAVELENGTH nm ABSORB ANCE Fi...

Page 7: ...SpO2 value For example if the 80 FingerSim was being used at 90 F ambient temperature then the expected simulation would be increased to 83 74 75 76 77 78 79 80 81 82 83 84 85 86 40 45 50 55 60 65 70...

Page 8: ...lder see Figure 9 Figure 9 Sensor Cable Routing Through Holder C Sensor Alignment and Position The light absorption characteristics of the FingerSim are specified for light passing directly through th...

Page 9: ...Pulse Rate readings WARNING FingerSim are fragile and must be handled with care they contain glass Figure 10 Positioning the FingerSim in the Holder B Attach the Sensor Attach the sensor under test to...

Page 10: ...ngs between manufacturers is possible due to each manufacturer s interpretation of how the SpO2 value relates to the absorption of the red and infrared light in blood see Theory of Operation Figure 12...

Page 11: ...c 20dB to 40dB Operating Temperature Range 65 F to 90 F Long Term Storage Temperature Range 32 F to 104 F Typical infrared percent modulation when squeezed 0 to 5 Red to Infrared Ratio a c 72 5 F and...

Page 12: ...the applicable warranty period i e before the calibration date This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any FingerSims or accessories delivered to the purch...

Page 13: ...or leaking FingerSim Replace FingerSim Outside calibration date Replace FingerSim Improper sensor attachment to the FingerSim Ensure the sensor is attached to the FingerSim per section VIII B Damaged...