BC Biomedical FINGERSIM Instruction And Service Manual Download Page 10

 

10 

C. Quiescent 

Test 

After attaching the sensor allow the oximeter system under test about 30 seconds to stabilize.  The oximeter 
system should recognize this as a no pulse condition. 

 
D. 

Oxygen Saturation Test 

Gently press the FingerSim’s™ color coded end rhythmically keeping a minimum pressure on the FingerSim™  
throughout the pulse cycle (see Figure 12).  Generate the pulse by slightly increasing and decreasing the 
pressure.  Do not tap or move your finger away from the FingerSim™.  This may create unwanted motion 
artifact by rocking the FingerSim™.   The oximeter system should recognize the simulated pulse and display a 
pulse rate correlating to the input.  The SpO

2

 display should approximate the nominal value of the FingerSim™.   

 

NOTE:

    Some slight variation in the SpO

2

 readings between manufacturers is possible due to each  

          manufacturer’s interpretation of how the SpO

2

 value relates to the absorption of the red and infrared   

          light in blood (see Theory of Operation). 

 

 

 

Figure 12.  Generating a Simulated Pulse 

 

E. 

Pulse Rate Test 

Vary the input pulse rate by rhythmically pressing the FingerSim™ colored coded end at fast and slow rates.  
The pulse oximeter system should indicate high and low pulse rates.  Verify that the pulse oximeter system 
recognizes each pulse generated and that there are no extra pulses indicated by the oximeter. 

 

 

F. 

Pulse Amplitude Test 

The amplitude of the pulse waveforms being generated can be varied between 0% and 5% modulation levels by 
changing the amount of pressure used to generate the pulse waveform.  These amplitude changes should be 
displayed by the oximeter on the perfusion indicator (various display techniques have been used by oximeter 
manufacturers: colored coded L.E.D.s, L.E.D. bar graphs, L.C.D. waveforms, etc.). 

Summary of Contents for FINGERSIM

Page 1: ...M PULSE OXIMETER TEST SYSTEM BC Biomedical BC Group International Inc 9415 Gentry Ave St Louis MO 63125 PROBLEMS Check Section XII Trouble Shooting Chart then if still troubled call BC Biomedical Cust...

Page 2: ...ICATIONS 11 A FingerSim 11 B Holder 11 IX SERVICE AND MAINTENANCE 12 X WARRANTY 12 XI ACCESSORIES 12 XII TROUBLE SHOOTING CHART 13 FIGURES Figure 1 FingerSim 4 Figure 2 Replacement Holder 4 Figure 3 8...

Page 3: ...possible operating conditions a pulse oximeter may encounter Use the FingerSim as an adjunct to other indications to determine proper pulse oximeter operation CAUTION Improper insertion of the Finger...

Page 4: ...od as measured by an oximeter at different oxygen saturation values nominally 97 90 and 80 The 97 90 and 80 FingerSims are easily identified by the color coded end caps Red 97 Blue 90 and Black 80 Squ...

Page 5: ...typical finger In addition the red and infrared photo spectrometric light absorption of the inner solution approximates arterial blood as seen by the oximeter at 80 90 and 97 oxygen saturation levels...

Page 6: ...RB ANCE Figure 3 80 SpO2 Absorption Spectrum Figure 4 90 SpO2 Absorption Spectrum Figure 5 97 SpO2 Absorption Spectrum FingerSim LIQUID 90 SpO2 NOMINAL 0 4 620 720 820 920 WAVELENGTH nm ABSORB ANCE Fi...

Page 7: ...SpO2 value For example if the 80 FingerSim was being used at 90 F ambient temperature then the expected simulation would be increased to 83 74 75 76 77 78 79 80 81 82 83 84 85 86 40 45 50 55 60 65 70...

Page 8: ...lder see Figure 9 Figure 9 Sensor Cable Routing Through Holder C Sensor Alignment and Position The light absorption characteristics of the FingerSim are specified for light passing directly through th...

Page 9: ...Pulse Rate readings WARNING FingerSim are fragile and must be handled with care they contain glass Figure 10 Positioning the FingerSim in the Holder B Attach the Sensor Attach the sensor under test to...

Page 10: ...ngs between manufacturers is possible due to each manufacturer s interpretation of how the SpO2 value relates to the absorption of the red and infrared light in blood see Theory of Operation Figure 12...

Page 11: ...c 20dB to 40dB Operating Temperature Range 65 F to 90 F Long Term Storage Temperature Range 32 F to 104 F Typical infrared percent modulation when squeezed 0 to 5 Red to Infrared Ratio a c 72 5 F and...

Page 12: ...the applicable warranty period i e before the calibration date This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any FingerSims or accessories delivered to the purch...

Page 13: ...or leaking FingerSim Replace FingerSim Outside calibration date Replace FingerSim Improper sensor attachment to the FingerSim Ensure the sensor is attached to the FingerSim per section VIII B Damaged...

Reviews: