ZERO ADJUST
(BLINKS)
( OFF)
SPAN ADJUST
(BLINKS)
(BLINKS)
SETPOINT
ENTRY
(OFF) (BLINKS)
FILTER SELECT
4.2.4. Exit from Setup & Calibration
To exit from setup and calibration mode, press the
key to save the changes and go to
the next step. Press the
and the
keys simultaneously to exit from the setup mode
and return to the operation mode.
STEPS LED’s POSITION DESCRIPTION
GREEN RED
Put the weight equal to SP1 or SP2 on load receptor
: If this key is pressed, the load is stored as SP1 value for O1
: If this key is pressed, the load is stored as SP2 value for O2
SETPOINT ENTRY
Off Flash
SPAN ADJUST
Flash Flash
Load your scale. The load value must be between 10% to 100% of your
maximum. Adjust the output value to the required value for your load by
using keys described below:
: Pressing this key will decrease the output value. Pressing this key
continuously will decrease output faster. While pressing this key
continuously, pressing key will decrease the output value very
high rate.
: Pressing this key will increase the output value. Pressing this key
continuously will increase output faster. While pressing this key
continuously, pressing key will increase the output value very
high rate.
.
High filtering ( Low frequency cut off / slower performance ) ( Default
) Press key to adjust the filter low filtering
Low filtering ( High frequency cut off / faster performance )
Press key to make the filter high filtering
Note: If it does not create you weighing problem, prefer high filtering
for increasing the stability of the weighing system.
FILTER SELECT
On
On
On Flash
ZERO ADJUST
Flash Off
: Pressing this key will decrease the output value. Pressing this key
continuously will decrease output faster. While pressing this key
continuously, pressing key will decrease the output value very
high rate.
: Pressing this key will increase the output value. Pressing this key
continuously will increase output faster. While pressing this key
continuously, pressing key will increase the output value very
high rate.
.
Unload the scale. Adjust the output value to 0 V or 4 mA by using keys
described below :