11
Scan Delay
If the injection system software version is 17A or higher, the scan delay must be programmed
from the scanner.
ISI Module Cables
Maintenance:
A periodic inspection of the ISI Interface Cables is a recommended to ensure
that there are no obvious defects, such as cuts or breaks in the cable, which could disrupt
signal integrity. Ensure that the connectors are properly seated.
NOTE:
Scanner specific cables are typically available from the scanner
manufacturer. Contact the scanner manufacturer for details.
Power Cable
Specifications
The specifications required by the MEDRAD
®
Stellant Imaging System Interface (ISI) 700
Module relative to the power cable (plug, receptacle, and cord) are:
•
Operating Temperature: 60°C minimum
•
Receptacle Type: IEC-60320 C13
•
Normal Cord Voltage: 300 VAC minimum
•
Wire Gauge: 1.00 mm
2
minimum
•
Cord Type: IEC 60245-1, Annex A, Designation 53, or IEC 60227-1,
Annex A, Designation 53 Certified
•
Cord Length: 3 m maximum
The power cable must meet applicable plug, cord, and receptacle specifications including type,
voltage, current, and safety approval markings for the country in which the power cable is being
used.
ISI Module System
Messages and
Descriptions
The text box messages appear on the injector screen.
Text Box Message
What it means
Actions to be taken
Display and Control Unit (DCU):
“Scanner not ready”
Certegra Workstation:
"Imaging system not ready"
When the ISI module is configured "On", or
when the user attempts to arm the injector,
the system checks for the signal, "Scanner
Ready." If the scanner is not ready, this
message results.
Ensure the scanner is ready. Check that all
cabling between the injector, ISI module,
and scanner are connected. Ensure power
indicator on ISI module is lit.
Display and Control Unit (DCU):
"Injector Disarmed by Scanner"
Certegra Workstation:
"Injector Disarmed by Scanner"
The injector was "Armed" or injecting, and
a disarm signal was received from the
scanner.
OR
The ISI module is not communicating with
the Injector and an attempt is made to
Arm.
In some cases, this can be normal
operation. However, if this message was
received in error, then check to ensure that
all cabling between the injector, ISI module,
and scanner are connected. Ensure that
the scanner is properly configured.
Display and Control Unit (DCU):
"ISI module not ready"
Certegra Workstation:
Not Applicable -- No message
This message appears when two events
occur simultaneously:
(1) The user is trying to configure the ISI
module, and
(2) The injector is not communicating with
the ISI module.
Check to ensure that all cabling between
the injector, ISI module, and scanner are
connected. Ensure power indicator on ISI
module is lit.
Summary of Contents for MEDRAD SCT 110
Page 2: ......
Page 4: ...MEDRAD Stellant ISI700ModuleOperationandInstallationManual ii...
Page 6: ...MEDRAD Stellant ISI 700 Module Operation and Installation Manual iv...
Page 22: ...MEDRAD Stellant ISI700ModuleOperationandInstallationManual 16...
Page 28: ...MEDRAD Stellant ISI700ModuleOperationandInstallationManual 22...
Page 29: ......