Signals and Descriptions

*The ISI scanner interface is a 37-pin D shell connector (J1), located on the ISI module. The 
following signals are provided to work properly with the scanner.

**The scanner manufacturer is responsible for assuring the scanner will not respond to a Scan 
Start signal until it is safe to do so.

Caution:

 System damage may occur if voltage is applied to J1 Pins 16, 18, 23, 25, 27, 

or 33. Only connect these pins to isolated switch or relay contacts in the scanner.

Signal

Operation

D

escription

*J1 Pin 

Numb

er

Voltage Logic

Scanner 

Read

y

This is an input signal fro

m

 the 

scanner 

to the injector. It 

tells the 

injector that the 

scanner is ready. This 

signal en

sures that

 the 

injecto

r cannot 

be armed unless the scanner 

is ready. 

This signal is used to pre

vent a 

condition where you

 could receive 

contrast 

media

 but no scan.

When the signal is high, it 

allows 

the injector 

to arm. If 

the 

user wants to arm the injector, 

and this signal is low, the 

injector cannot be armed.

25

If 

+5v is switched from pin

 16 

to 25 by the scanner, the 

injector can be a

rmed. If pin 

25 is zero volts, the injector 

cannot be armed.

Injector 

Di

sa

rm

This is an input signal fro

m

 the 

scanner 

to the injector. It 

allows the 

scanner 

to stop the 

injection at

 any 

time by 

disarming it. This 

signal would 

immediately st

op the injection of 

contrast.

If the in

jector is armed (this 

signal would be high), making 

this sig

nal go to zero disarms 

the injector. 

25

Note that p

in 25 performs a

 

dual function. If it is h

igh 

(+5v) the injector is allowed 

to arm. If it

 goes low (0v), it 

will disarm the injector.

Start

 Injector 

Requ

est

This is an input signal fro

m

 the 

scanner 

to the injector. It 

allows the 

scanner 

to request that the injector 

start injecting. If the injector is armed, 

and

 the scanner initiates a "Start 

Request", fluid will be delivered, 

oth

erwise the signa

l is ignored.

The signal from the scanner 

is 

normally low. When it goes 

high, th

e "Start Request" is sen

to the scanner.

27

Pin 2

7 is normally at 0v. 

When the scanner switches 

the isolated 5v 

from pin 16 to 

pin 27, 

the "S

ta

rt Request" 

is 

made.

Injector 

Read

y

This is an output sig

nal from the 

injector to the scanner indicating the

 

injector is armed or 

holding. 

When the 

scanner 

is initiating the entire 

scanning sequence, this signal is used 

to 

notify the 

scanner that the injector is 

ready to accept a start or has entered 

a "hold" phase. The injector will start 

when receiving a "Start injector 

request."

An injector Ready 

signal is sent 

to the 

scanner once 

the injecto

is armed.

1 &

 2

The scanner provides a 

signal on pin 2 of J1. When 

the injector is armed, this 

signal is switched back to 

the scanne

r through pin 1. 

The signal is low while 

inj

ecting.

Summary of Contents for MEDRAD SCT 110

Page 1: ...CT Injection System CT Injection System Operation and Installation Manual MEDRAD Stellant Imaging System Interface ISI 700 Module...

Page 2: ......

Page 3: ...as well as required calibration s that are needed Required reading includes the instructions for use and other materials provided with the device This also includes any hardware and software updates...

Page 4: ...MEDRAD Stellant ISI700ModuleOperationandInstallationManual ii...

Page 5: ...nstallation 6 Operational Checkout 8 Configuration 9 Overview of Operation 9 Injector Head Indicator Lights 10 Details of Operation 10 Hold and Test Inject features 10 Scan Delay 11 ISI Module Cables...

Page 6: ...MEDRAD Stellant ISI 700 Module Operation and Installation Manual iv...

Page 7: ...f allowing an injector to interface with a CT scanner Contraindications The ISI module is not to be used for any other use for which the device is not indicated This device is not to be used for drug...

Page 8: ...mains the same Throughout the manual Display and Control Unit DCU and Certegra Workstation will be used to refer to the two when there are differences in how information is presented on the screens Sy...

Page 9: ...in each carton ISO 15223 1 5 4 4 Warning Indicates hazardous voltages ISO 7010 W012 Indicates alternating current IEC 60417 5032 Indicates scanner connection Identifies a terminal suitable for direct...

Page 10: ...communicating to the ISI module The scanner signals are in a state that will allow arming the injector The user is not able to arm the injector The injector is configured for ISI and it is communicat...

Page 11: ...use it to fail Installation must be completed by properly trained personnel Contact your scanner manufacturer for configuration and installation assistance Patient injury could result from using impro...

Page 12: ...a soft cloth or paper towel dampened with cleaning solution Do not use strong cleaning agents and solvents Warm water and a mild disinfectant are all that are required to clean the injector Do not use...

Page 13: ...Connect the power cable to the ISI module Ensure power indicator is illuminated 5 If a Certegra Workstation is being installed as an upgrade to an existing injector with an ISI 700 option determine w...

Page 14: ...Operational Checkout A basic functional checkout of the ISI module should be included as part of regular maintenance Verifying proper operation of the ISI module will help in detection of any problems...

Page 15: ...itiation The ISI module supports the following two methods 1 By pressing the injector start switch the injector notifies the scanner that it can begin the scanning sequence 2 The scanner initiates the...

Page 16: ...d with a short flash long flash pattern Details of Operation 1 Program the scanner You will see the ISI Ready icon when the scanner is ready 2 Program the injector 3 Arm the injector NOTE For specific...

Page 17: ...injector screen Text Box Message What it means Actions to be taken Display and Control Unit DCU Scanner not ready Certegra Workstation Imaging system not ready When the ISI module is configured On or...

Page 18: ...ntact Bayer for further support Schematic Diagram Note Maximum rating for scanner relay contacts 0 75 A 30 VDC Minimum rating for scanner relay contacts 10 a 10mV Display and Control Unit DCU ISI modu...

Page 19: ...at any time by disarming it This signal would immediately stop the injection of contrast If the injector is armed this signal would be high making this signal go to zero disarms the injector 25 Note...

Page 20: ...for a Class 1 Medical Device EMI RFI The injection system is classified as Group 1 Class A equipment per the requirements of IEC 60601 1 2 Accessories provided by Bayer will also comply with this sta...

Page 21: ...Mode of Operation Per IEC 60601 1 the mode of operation for the ISI module is continuous It is capable of operation under normal load for an unlimited period without excessive temperature being develo...

Page 22: ...MEDRAD Stellant ISI700ModuleOperationandInstallationManual 16...

Page 23: ...ing and measurement techniques Surge immunity test IEC 61000 4 6 Electromagnetic compatibility EMC Part 4 6 Testing and measurement techniques Immunity to conducted disturbances induced by radio frequ...

Page 24: ...rence by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the system as recommended below according to the maximum output power of the communicat...

Page 25: ...use in the electromagnetic environment specified below The customer or user of the system should assure that it is used in such an environment Immunity test IEC 60601 Test Compliance Level Electromag...

Page 26: ...d with the following symbol Non ionizing Radiation Symbol IEC TR 60878 5140 NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Elec...

Page 27: ...ctor Head Indicator Lights 10 Installation 6 J L J1 Pins 13 M O Maintenance 11 Messages and Descriptions 11 Operation Details of 10 Operation Overview 9 Operational Checkout 8 P R Rating maximum 12 Ra...

Page 28: ...MEDRAD Stellant ISI700ModuleOperationandInstallationManual 22...

Page 29: ......

Page 30: ...in this publication at any time without prior notice or obligation Please contact your authorized Bayer representative for the most current information All patient data that appear in this document ar...

Reviews: