Preventive Maintenance
5
SIGMA Spectrum Infusion System
Pump Operating Software Version 6.02 and 6.05
Preventive Maintenance
For Use With MDL Version 6
Pump Case
Pump Display
Keypad
Keyhole
Pump Door
IV Set Tube Channel
Pole Clamp
External Power Adaptor/Accessory Connection port
NOTE:
The IV tube channel must be free of foreign materials and dry for both testing and
use.
2. Verify presence and legibility of all labels. Replace/order labels as necessary.
For label location, see Appendix A: “Technical Drawings” in the Service Manual.
Company ID label
Power/Bolus label
Certification label
Serial Number label
Tubing ID label
Direction of Flow label
Replace/order labels as necessary.
3. Record the findings on the Preventive Maintenance Check Sheet.
Visual Inspection - AC Power Adaptor
To inspect the AC Power Adaptor:
1. Inspect the AC Power Adaptor for visual evidence of damage and/or defects such as cracks,
bent prongs, cuts or exposed wires or dents.
AC Power Adaptor Module
AC Power Adaptor Cord
AC Power Cord Adaptor Retainer
If the AC Power Cord Adaptor Retainer is damaged,
Remove the Retainer and inspect the 6-pin Plug Connector (on the AC Power Adaptor
cord).
Inspect the External Power Adaptor/Accessory Connection port (on the back of the Pump)
for signs of damage such as cracks or missing pins.
NOTE:
If damage is found on the AC Power Adaptor, it may be sent for service repair
(without the Pump).
2. If no signs of damage are found, plug the AC Power Adaptor into a working wall receptacle.