18
Preventive Maintenance
Manual 41394
SIGMA Spectrum Infusion System
Revision A
Preventive Maintenance
18. Press RUN to resume infusion.
The Check Flow screen appears.
19. Press the
yes
soft key to proceed to the Infusion Running screen.
20. Press
RUN/STOP
to stop the infusion.
NOTE:
In the event of a test failure, confirm failure by repeating the test with a new
(unused) IV set.
21. Record the findings on the Preventive Maintenance Check Sheet.
22. If performing additional tests, ensure to re-prime the IV set to remove any remaining air in line.
Battery Capacity Test
Perform the Battery Capacity Test if a problem is suspected with the Pump’s Battery Module. For
instance, if the charge does not last as long as it should.
NOTE:
Prior to starting the test, charge a Standard Battery for a minimum of 12 hours. Charge a
Wireless Battery Module for a minimum of 16 hours.
To test the battery capacity:
1. Plug the AC Power Adaptor into a working wall outlet and verify green LED is lit while plugged
in.
2. Adjust an IV test set in a loop configuration, so as to circulate fluid around the loop.
3. Press
ON/OFF
to power on the Pump
4. Load the IV set into the Pump following the display prompts. Verify that the IV set is properly
loaded.
NOTE:
If the IV set is not properly loaded, the test results will be invalid.
5. Press the
options menu
soft key.
6. Select
User Options
and press
OK
.
7. Select
Display Settings
and press
OK
.
8. Select
Display Adjust
and press
OK
.
9. Set the Backlight Level to
10
and press
OK
.
10. Press
exit
twice to return to the Options screen.
11. Select the
Biomed Options
and enter the first level Biomed password.
12. Select
Pump Testing
and press
OK
.
13. Select
Battery Test
and press
OK
.
14. Select
Battery Life Test
and press
OK
.