DTA 100 C
User-defined measurement
822-129-8
55 / 90
11.1.4 Setting the stirring time during the hold time
Recommendation:
As the homogeneity of the oil sample has a strong influence on the
quality of the measurement results, stir the oil sample during the measurement process.
Main menu
>
User-defined measurement
>
...
>
Stirring in hold time
1. Select menu item
Stirring in hold time
.
2. Select menu item
Change
.
3. Enter the stirring time during the hold time before the first measurement using the
number keys on the membrane keyboard.
4. Click
Save
to confirm the entry.
11.1.5 Setting the slew rate for the test voltage
Main menu
>
User-defined measurement
>
...
>
Volt. rise
1. Select the menu item by pressing the arrow keys.
2. Select menu item
Change
.
3. Enter the slew rate for the test voltage with the number keys on the membrane keypad.
4. Click
Save
to confirm the entry.