Device settings
DTA 100 C
40 / 90
822-129-8
8.6.2
Set the hold time before the first measurement
Main menu
>
Device settings
>
...
>
Hold time
1. In the main menu, select the menu item
Device settings
.
2. Select the menu item
Next
to confirm the selection.
3. In the
Device settings
menu select the menu item
ASTM settings
and then menu item
Next
.
4. Select menu item
Hold time
.
5. Select menu item
Change
.
6. Choose one of the following settings by pressing the arrow keys:
3 min:
The hold time is 3 minutes.
30...60 min:
The hold time can be set to between 30 and 60 minutes.
7. If you have enabled the
30...60 min
setting, select the menu item
Next
.
8. Enter the desired hold time.
When testing ester liquids in accordance with ASTM D1816:2012 the hold time before
the first measurement must be at least 30 mins.
9. Click
Save
to confirm the entry.